Toxic levels of acetaminophen produce a major positive interference on Glucometer Elite and Accu-chek Advantage glucose meters.

نویسندگان

  • L J Cartier
  • P Leclerc
  • M Pouliot
  • L Nadeau
  • G Turcotte
  • B Fruteau-de-Laclos
چکیده

ied this possibility by performing the enzymatic assay on controls containing increasing concentrations of ACET in the presence of different therapeutic concentrations of NAC (Fig. 1A). Evidently, NAC concentrations $100 mg/L cause a marked negative bias, reaching total suppression of the reaction in the presence of 10 g/L NAC. In contrast, similar concentrations of NAC have no effect on the ACET TDx immunoassay (Fig. 1B). According to the TDx method manual, concentrations of up to 1000 mg/L NAC cause only a minimal bias of ,1 mg/L ACET in this assay. Plasma concentrations of NAC after administration of the common initial dose of 150 mg/kg of body weight over a period of 15 min have been reported to range from 304 to 875 mg/L (5). NAC concentrations in this range, and certainly the higher concentrations of blood NAC reached after larger dosages during infusion, markedly impede ACET determination by the enzymatic assay (Fig. 1A). That the drug of choice for treating ACET intoxication produces a very large negative bias in a commonly used enzymatic-colorimetric assay of ACET is worth noting. On the basis of the present observation, we recommend that patients’ blood should be sampled for ACET before the initiation of NAC treatment. Alternatively, enzyme-based colorimetric methods should be replaced by immunoassays or HPLC assays of ACET in which NAC causes no interference.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

مقایسه قدرت دستگاه گلوکومتر Accu-chek با Elite در تشخیص دیابت با استفاده از خونریزی متعاقب Probing

Abstract Background: Recent studies have assessed the validity of glucometer devices in estimating the glucose level of gingival blood during routine periodontal examination (Bleeding On Probing). The purpose of this study was to evaluate whether blood oozing from gingival tissues during routine periodontal examination can be used for determining g...

متن کامل

Effect of fluoride-containing tubes on accuracy of glucometers.

G lucometers are used to quickly obtain glucose concentrations in the hospital and at home. For quality assurance, the American Diabetes Association recommends periodically comparing the results from glucometers with those from referenced laboratories (1). Usually, measurement using glucometers is done one-by-one by technicians, which is costly regarding manpower. In contrast, measurement using...

متن کامل

Dynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices.

BACKGROUND It has been demonstrated that dynamic electrochemistry can be used to correct blood glucose measurement results for potentially interfering conditions, such as humidity, hematocrit (HCT) variations, and ascorbic acid. The purpose of this laboratory investigation was to assess the potential influence of hematocrit variations on a variety of blood glucose meters applying different meas...

متن کامل

Assessment of three frequently used blood glucose monitoring devices in clinical routine.

BACKGROUND Self-monitoring of blood glucose plays an important role in the management of diabetes and has been shown to improve metabolic control. The use of blood glucose meters in clinical practice requires sufficient reliability to allow adequate treatment. Direct comparison of different blood glucose meters in clinical practice, independent of the manufactures is scarce. We, therefore, aime...

متن کامل

Monitoring neonatal hypoglycemia with the Accu-chek advantage II glucose meter: the cautionary tale of galactosemia.

To the Editor: Hand-held glucose meters are a practical solution for measuring glucose in neonatal intensive care units and special care nurseries because of their portability, the immediacy of results, and minimal blood volume requirements compared with dispatching samples for central laboratory analysis. We have used the Roche Accu-chek Advantage II meter for some time in our neonatal intensi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 44 4  شماره 

صفحات  -

تاریخ انتشار 1998