Determination of the dead time of a stopped-flow fluorometer.

نویسندگان

  • P Brissette
  • D P Ballou
  • V Massey
چکیده

This investigation was carried out to develop a convenient alternative method for examining the performance and determining the dead time of a stopped-flow fluorometer. We examined the kinetics for the formation of the fluorescent Mg2+-8-hydroxyquinoline chelate in aqueous solutions. The reversible association of the Mg2+ ion with 8-hydroxyquinoline is a second-order process whose on and off rate constants are dependent on pH. We estimated that the Mg2+ ion chelate has a fluorescence quantum yield of 0.02 in aqueous solutions. Using this reaction we measured the dead time of a stopped-flow fluorometer at different pH values. Measurements of the dead time were found to be reproducible and accurate. The Mg2+-8-hydroxyquinoline reaction fulfills the requirements for a convenient test reaction for dead time measurement of stopped-flow fluorometers. Although the usefulness of the reaction is primarily to determine the dead times of stopped-flow instruments operating in the fluorescence mode, the reaction can also be used for testing an instrument operating in the absorbance mode.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Flow-Injection Amperometric Determination of Ascorbic Acid Using a Graphite-Epoxy Composite Electrode Modified with Cobalt Phthalocyanine

A flow injection method is reported for the determination of ascorbic acid based on amperometric detection, using a cobalt(II) phthalocyanine (CoPc) modified graphite-epoxy electrode. The amperometric respons was evaluated with regard to pH, ionic strength of the electrolyte, flow rate of the carrier solution, injected sample volume and conditioning time of the electrode. The limit of detec...

متن کامل

FLOW INJECTIOW DETERMINATION OF FLUORIDE BY MURTN-ALUMINUM REAGEM WITH SPECTROEUORIMIETRIC DETECTION

A flow injection system is reported for the determination of fluoride based on its effects on the fluorescence intensity of Al-morin complex at pH 3.0 in ethanol-warn mixture. The effect of important parameters such as excitation and emission wavelengths, pH, reagents concenkations, flow rate of the carrier stream, sample injection volume, time of measurepent and interferences are reported. ...

متن کامل

Statistical model of daily flow in arid and semi arid regions

Abstract Stochastic model for synthetic data generation is not available for daily flows of intermittent streams. Such a model is required in the planning and operation of structures on an intermittent stream for purposes where short time flow fluctuations are important. In this study a model is developed for such a case. The model consists of four steps: determination of the days on which fl...

متن کامل

Calibration of stopped-flow spectrophotometers using a two-step disulfide exchange reaction.

A coupled two-step reaction of Ellman’s reagent (5,5’-dithiobis(2-nitrobenzoic acid)) with excess thioglycerol produces a progress curve composed of two superimposed exponentials. The ratio of the two pseudo-first-order rate constants equals 22.5 and does not vary appreciably with changes of either pH value or temperature. Because the ratio of the two amplitudes is defined by the ratio of the r...

متن کامل

مدلسازی عددی حوضچه های بنادر رودخانه‌ای و بررسی تاثیر رژیم جریان بر گردابه‌های افقی

Using dead zones of rivers for berthing purpose may involve some problems of which the most serious are sedimentation and silting. These processes depend on water flow pattern in these zones. In this paper, a two dimensional depth averaged model is explained for flow modeling in river dead zones. Governing equations of fluid flow are discretised using the “Control Volume” method and...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Analytical biochemistry

دوره 181 2  شماره 

صفحات  -