Dynamic fatigue performance of implant-abutment assemblies with different tightening torque values.

نویسندگان

  • Dandan Xia
  • Hong Lin
  • Shenpo Yuan
  • Wei Bai
  • Gang Zheng
چکیده

Implant-abutment assemblies are usually subject to long-term cyclic loading. To evaluate the dynamic fatigue performance of implant-abutment assemblies with different tightening torque values, thirty implant-abutment assemblies (Zimmer Dental, Carlsbad, CA, USA) were randomly assigned to three tightening groups (24 Ncm; 30 Ncm; 36 Ncm), each consisted of 10 implants. Five specimens from each group were unscrewed, and their reverse torque values recorded. The remaining specimens were subjected to a load between 30 N~300 N at a loading frequency of 15 Hz for 5 × 10(6) cycles. After fatigue tests, residual reverse torque values were recorded if available. In the 24 Ncm tightening group, all the implants fractured at the first outer thread of the implant after fatigue loading, with fatigue crack propagation at the fractured surface showed by SEM observation. For the 30 Ncm and 36 Ncm tightening groups, a statistical significant difference (p<0.05) between the unloaded and loaded groups was revealed. Compared with the unloaded specimens, the specimens went through fatigue loading had decreased reverse torque values. It was demonstrated that insufficient torque will lead to poor fatigue performance of dental implant-abutment assemblies and abutment screws should be tightened to the torque recommended by the manufacturer. It was also concluded that fatigue loading would lead to preload loss.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

بررسی میزان کاهش تورک در اباتمنت‌های کست شونده با آلیاژ کروم کبالت پس از اعمال بارگذاری دوره‌ای

Background and Aims: For the clinical success of implant-retained prostheses, fitness of matching components and stability of the applied torque to the implant-abutment assembly under dynamic occlusal loading is essential. The aim of this study was to evaluate the torque loss in Co-Cr castable abutments after cyclic loading. Materials and Methods: 10 castable abutments were cast in Co-Cr alloy...

متن کامل

In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening

Preload is applied to screws manually or using a torque wrench in dental implant systems, and the preload applied must be appropriate for the purpose. The aim of this study was to assess screw loosening and bending/torsional moments applied by clinicians of various specialties following application of manual tightening torque to combinations of implants and abutments. Ten-millimeter implants of...

متن کامل

Comparison of the Effect of Two Different Abutment Diameters on Detorque Value Using Cyclic Loading and Thermocycling

Background and Aim: Screw loosening is a common problem with both screw-retained and cemented implant restorations. It is assumed that the abutment diameter affects detorque value and screw loosening. We aimed to determine the effect of two different abutment diameters on detorque value using cyclic loading and thermocycling. Materials and Methods: This in-vitro experimental study was conducte...

متن کامل

Manual maximum closing torque in three implant abutment systems

Background and Aims: Diferences whithin shape and size of hand wrenches in diferent systems could affect the torque applied. Therefore, evaluation of wrenches has been emphasized. The aim of the present study was to compare the torque values of the long and short torque wrenches in different implant systems including ITI, Implantium and Nobel Biocare. Materials and Methods: In this experimenta...

متن کامل

تحلیل بیومکانیکی پیچ و مخروط و آنالیز حساسیت شل شدن اباتمنت در ایمپلنت‌های دندانی

Background and Aims: Different mechanisms have been developed for connecting abutment to implant. One of the most popular mechanisms is Tapered Integrated Screw (TIS), which is a Tapered Interference Fit (TIF) with a screw integrated at the bottom of that. The aim of this study was to investigate the mechanism of TIS and effective factors in employing TIS during design and implementation proces...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Bio-medical materials and engineering

دوره 24 6  شماره 

صفحات  -

تاریخ انتشار 2014