Root canal preparation with a novel nickel-titanium instrument evaluated with micro-computed tomography: canal surface preparation over time.

نویسندگان

  • Ove A Peters
  • Claudia Boessler
  • Frank Paqué
چکیده

INTRODUCTION The aim of this part of an ongoing study was to describe the dentin removal ability of a novel nickel-titanium instrument, the self-adjusting file (SAF), by using micro-computed tomography. METHODS Twenty maxillary incisors were scanned preoperatively at 20-microm resolution and postoperatively after up to 6 minutes of preparation with an SAF with 1.5-mm or 2-mm diameter. SAFs were operated with continuous irrigation in a handpiece that provided an oscillating, in-and-out movement. Changes in canal volume compared with preoperative values as well as unprepared canal surface area were determined. Data were normally distributed and compared by analysis of variance and regression analyses. RESULTS Preoperatively canal volumes were statistically similar in both groups (9.86 +/- 3.97 mm3 and 9.80 +/- 2.67 mm3). Volumes increased during preparation to 13.58 +/- 3.85 mm3 (after 6 minutes with SAF 1.5 mm) and 16.43 +/- 3.64 mm3 (after 5 minutes with SAF 2.0 mm), and overall canal shapes were adequate. Unprepared canal surface area decreased from 63.0% +/- 15.1% (2 minutes with SAF 1.5 mm) to 8.6% +/- 4.1% (5 minutes with SAF 2.0 mm). CONCLUSIONS Preparation of straight root canals in maxillary anterior teeth left little canal surface uninstrumented after shaping with the SAF. The timeframe of clinical application will depend on the amount of desired dentin removal and done with an SAF selected on the basis of apical gauging.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Micro-computed tomography evaluation of the preparation of long oval root canals in mandibular molars with the self-adjusting file.

INTRODUCTION The aim of this study was to assess the shaping potential of a novel nickel-titanium instrument, the self-adjusting file (SAF), in long oval root canals in distal roots in mandibular molars. METHODS Twenty mandibular molars with long oval distal root canals were selected and scanned preoperatively and postoperatively by using micro-computed tomography at an original resolution of...

متن کامل

Re: A micro-computed tomographic evaluation of apical root canal preparation using three instrumentation techniques.

AIM To investigate the morphological changes in the apical third of the root canal after preparation with three techniques. METHODOLOGY Forty molar teeth were scanned using micro-computed tomography before and after instrumentation with: Group 1 - stainless steel K-files using the balanced force technique; Group 2 - stainless steel K-files (balanced force) and then refining the apical prepara...

متن کامل

ProTaper rotary root canal preparation: assessment of torque and force in relation to canal anatomy.

AIM To investigate physical parameters for ProTaper nickel-titanium (NiTi) rotary instruments whilst preparing curved canals in maxillary molars in vitro. METHODOLOGY A novel torque-testing platform was used to prepare root canals in 15 extracted human maxillary molars with ProTaper rotary instruments. Peak torque and force was registered along with numbers of rotations required to shape the ...

متن کامل

Effects of root canal preparation on apical geometry assessed by micro-computed tomography.

INTRODUCTION Previous micro-computed tomography analyses of root canal preparation provided data that were usually averaged over canal length. The aim of this study was to compare preparation effects on apical root canal geometry. METHODS Sixty extracted maxillary molars (180 canals) used in prior studies were reevaluated for analyses of the apical 4 mm. Teeth were scanned by using micro-comp...

متن کامل

ProTaper rotary root canal preparation: effects of canal anatomy on final shape analysed by micro CT.

AIM To evaluate the relative performance of ProTaper nickel-titanium (Ni-Ti) instruments shaping root canals of varying preoperative canal geometry. METHODOLOGY Extracted human maxillary molars were scanned, before and after shaping with ProTaper, employing micro computed tomography (muCT) at a resolution of 36 mum. Canals were three-dimensionally reconstructed and evaluated for volume, surfa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of endodontics

دوره 36 6  شماره 

صفحات  -

تاریخ انتشار 2010