Double left ventricular pacing following accidental malpositioning of the right ventricular electrode during implantation of a cardiac resynchronization therapy device
نویسندگان
چکیده
Accidental malpositioning of a right ventricular (RV) electrode has not been previously reported in the context of cardiac resynchronization therapy (CRT).The case of a 75-year old male patient with dilative cardiomyopathy, left ventricular (LV) ejection fraction 23%, New York Heart Association functional heart failure status stage III, left bundle branch block (LBBB) with QRS width of 136 ms, and misplacement of the RV lead to the LV apex during implantation of a CRT defibrillator is described.Following unremarkable implantation, routine interrogation of the CRT device on the first day after the implantation revealed uneventful technical findings. The 12-lead surface electrocardiogram (ECG) showed biventricular stimulation featuring a narrow QRS complex with incomplete right bundle branch block (RBBB) and R>S in V1. The biplane postoperative chest X-ray was graded normal. On routine follow-up one month later, a transthoracic echocardiogram revealed an increased ejection fraction of 51% but the RV lead was placed in the LV apex. An additional transesophageal echocardiogram exhibited an Eustachian valve guiding the lead via the patent foramen ovale through the mitral valve into the LV apex. Operative revision was scheduled and the active fixation lead was uneventful removed from the LV. A new electrode was inserted and placed in the RV apex.Accidental malplacement of the RV electrode to the LV may be difficult to diagnose in the context of CRT patients as a stimulated biventricular ECG with incomplete RBBB appearance is expected in this situation. Careful analysis of lateral radiographic views during the operation is important to ensure correct lead positioning. As timely revision is the preferred procedure, early routine transthoracic echocardiography may be considered for detection of malplacement.
منابع مشابه
Bifocal pacing in the right ventricle: an alternative to resynchronization when left ventricular access is not possible in end-stage heart failure patients.
BACKGROUND It has been reported that bifocal pacing (BiF) in the right ventricle might be an alternative to unsuccessful left ventricular lead implantation. This case report presents an assessment of the clinical and hemodynamic parameters during a three month follow-up in patients implanted with right ventricular BiF. METHODS Eight patients who underwent unsuccessful left ventricular lead im...
متن کاملCardiac Resynchronization Therapy: Lead Positioning and Technical Advances
Cardiac resynchronization therapy (CRT) is a therapeutic option for heart failure patients with a severely reduced left ventricular ejection fraction and left bundle branch block (Cleland et al., 2001). Ventricular resynchronization is achieved by biventricular pacing, usually via electrodes in the right ventricular apex and a left ventricular (LV) electrode positioned in a coronary vein. About...
متن کامل[Double-wire technique for implanting a left ventricular venous lead in patients with complicated coronary venous anatomy].
INTRODUCTION AND OBJECTIVES Occasionally, implanting a left ventricular pacing electrode for cardiac resynchronization therapy via the coronary sinus may be complicated by the presence of anatomical structures that obstruct the access to the target vein. Our objective was to report on experience using a double-wire technique for implanting left ventricular venous leads gained at the Dr Negrín H...
متن کاملSimultaneous His Bundle and Left Ventricular Pacing for Optimal Cardiac Resynchronization Therapy Delivery: Acute Hemodynamic Assessment by Pressure-Volume Loops.
BACKGROUND Previous studies have investigated the role of intrinsic conduction in optimizing cardiac resynchronization therapy. We investigated the role of fusing pacing-induced activation and intrinsic conduction in cardiac resynchronization therapy by evaluating the acute hemodynamic effects of simultaneous His-bundle (HIS) and left ventricular (LV) pacing. METHODS AND RESULTS We studied 11...
متن کاملTwiddling in cardiac resynchronization therapy: 'when length matters'.
We report on a 75-year-old woman who successfully underwent cardiac resynchronization therapy. Two weeks after implantation, right continuous pectoral stimulation appeared. Retraction and dislodgement of the right atrial and right ventricular leads as well as coiling of the three leads around the device generator were observed on chest radiography, but, surprisingly, the left ventricular lead r...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 8 شماره
صفحات -
تاریخ انتشار 2013