Analytical characterization of electrochemical biosensor test strips for measurement of glucose in low-volume interstitial fluid samples.

نویسندگان

  • M E Collison
  • P J Stout
  • T S Glushko
  • K N Pokela
  • D J Mullins-Hirte
  • J R Racchini
  • M A Walter
  • S P Mecca
  • J Rundquist
  • J J Allen
  • M E Hilgers
  • T B Hoegh
چکیده

BACKGROUND Minimally invasive interstitial fluid (ISF) sampling and glucose measurement technologies were integrated into a hand-held device for diabetic glucose monitoring investigations. METHODS Conventional electrochemical test strip technology (Bayer Glucometer Elite) was adapted to measure glucose in small (0.5-2.0 microL) samples of ISF. Test strip glucose measurements were performed on a commercial potentiostat and were compared to various reference glucose methodologies (YSI 2300 analyzer, microhexokinase procedure, Bayer Glucometer Elite). Characterizations of the integrated ISF sampling-glucose test strip design included accuracy and precision in various sample media (saline, ISF surrogates, diabetic ISF samples), sample volume dependence, test strip sterilization studies (electron beam, gamma irradiation), and diabetic ISF sampling and glucose measurements. RESULTS Glucose measurements were free from significant media effects. Sample volume variations (0.6-3.2 microL) revealed only modest dependence of glucose measurement bias on sample volume (-1.5% per microliter). Sterilization treatments had only a minor impact on glucose response and test strip aging and no significant impact on interferent responses of the glucose test strips. Diabetic subject testing under minimum fasting conditions of at least 2 h with integrated ISF sampling and glucose measurement gave low ISF glucose measurement imprecision (CV, 4%) and mean glucose results that were indistinguishable from reference (microhexokinase) ISF glucose measurements and from capillary blood glucose measurements (Glucometer Elite). CONCLUSIONS Conventional single-use, electrochemical glucose test strip and ISF collection technologies can be readily integrated to provide real-time ISF sampling and glucose measurements for diabetic monitoring applications.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Facile Synthesis and Electrochemical Performance of Graphene-Modified Cu2O Nanocomposite for Use in Enzyme-Free Glucose Biosensor

Graphene-modified Cu2O nanocomposite was synthesized under facile microwave irradiation of an aqueous solution and has been investigated as an enzyme-free glucose biosensor. Morphology and crystal structure of the graphene-modified Cu2O nanocomposite were investigated by using electron microscopy and X-Ray Diffraction (XRD) analyses. Also, the electrochemical performan...

متن کامل

The Inkjet Printing of Reducible AgNPs amperometric glucose biosensor Electrodes

The enzymes immobilization of the is crucially effective factor in biosensor preparation. Metal nanoparticles potentially able to immobilize the enzymes according to unique properties including large surface-to-volume ratio, high surface reaction activity, high catalytic efficiency, and strong adsorption ability. A novel and highly sensitive amperometric glucose biosensor was obtained by using ...

متن کامل

DESIGN AND FABRICATION OF ELECTROCHEMICAL BIOSENSOR BASED ON NANO-GRAPHENE TO DETECT SEROTONIN IN DIABETIC ZEBRA FISH

Background: A novel nanocomposite-modified electrode based on reduced graphene oxide (rGO) decorated with crown-ether and gold nanoparticles (GNPs) on the surface of a glassy carbon electrode (GCE) was fabricated to investigate 5-HT determination. Methods: The morphology of nanocomposite was characterized by scaning electron microscopy (SEM). Diabetic zebrafish was obtained by overfeeding via ...

متن کامل

The obtain optimum production conditions for Glucose Oxidase biosensor using software Qualtek-4

The Glucose Oxidase (Gox) electrochemical sensors were used to detect a little amount of glucose. In this paper, the pure platinum plate (length of 5 cm and a width of 3 mm), the Ag/AgCl electrode and platinum electrode (diameter of 1 mm) were applied as a working electrode (W.E), reference electrode and auxiliary electrode sequential. Factors measured, glucose concentration and pH is Phosphate...

متن کامل

The obtain optimum production conditions for Glucose Oxidase biosensor using software Qualtek-4

The Glucose Oxidase (Gox) electrochemical sensors were used to detect a little amount of glucose. In this paper, the pure platinum plate (length of 5 cm and a width of 3 mm), the Ag/AgCl electrode and platinum electrode (diameter of 1 mm) were applied as a working electrode (W.E), reference electrode and auxiliary electrode sequential. Factors measured, glucose concentration and pH is Phosphate...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 45 9  شماره 

صفحات  -

تاریخ انتشار 1999