Hysterosalpingography using a flat panel unit: Evaluation and optimization of ovarian radiation dose.
نویسندگان
چکیده
PURPOSE The aim of the present study was the evaluation and optimization of radiation dose to the ovaries (D) in hysterosalpingography (HSG). METHODS The study included a phantom study and a clinical one. In the phantom study, we evaluated imaging results for different geometrical setups and irradiation conditions. In the clinical study, 34 women were assigned into three different fluoroscopy modes andD was estimated with direct cervical TLD measurements. RESULTS In the phantom study, we used a source-to-image-distance (SID) of 110 cm and a field diagonal of 48 cm, and thus decreased air KERMA rate (KR) by 19% and 70%, respectively, for beam filtration: 4 mm Al and 0.9 mm Cu (Low dose). The least radiation exposure was accomplished by using the 3.75 pps fluoroscopy mode in conjunction with beam filtration: Low dose. In the clinical study,D normalized to 50 s of fluoroscopy time with a 3.75 pps fluoroscopy mode reached a value of 0.45 ± 0.04 mGy. Observers' evaluation of diagnostic image quality did not significantly differ for the three different modes of acquisition that were compared. CONCLUSIONS Digital spot radiographs could be omitted in modern flat panel systems during HSG. Fluoroscopy image acquisitions in a modern flat panel unit at 3.75 pps and a beam filtration of 4 mm Al and 0.9 mm Cu demonstrate acceptable image quality with an averageD equal to 0.45 mGy. This value is lower compared to the studied literature. For these reasons, the proposed method may be recommended for routine HSG examination in order to limit radiation exposure to the ovaries.
منابع مشابه
Assessment of Patient Absorbed Radiation Dose during Hysterosalpingography: A Pilot Study in Southwest Nigeria
Background: Hysterosalpingography (HSG) is an indispensable tool for diagnosing infertility in females. The procedure exposes female reproductive organs to ionizing radiation as the genitals are irradiated during the process. Investigating patient absorbed dose during the procedures is essential for effective radiological protection of the patient. Objective: This study aims to investigat...
متن کاملComparing Ovarian Radiation Doses in Flat-Panel and Conventional Angiography During Uterine Artery Embolization: A Randomized Clinical Trial
BACKGROUND Uterine artery embolization (UAE) is a minimally invasive procedure performed under fluoroscopy for the treatment of uterine fibroids and accompanied by radiation exposure. OBJECTIVES To compare ovarian radiation doses during uterine artery embolization (UAE) in patients using conventional digital subtraction angiography (DSA) with those using digital flat-panel technology. PATIE...
متن کاملIn vivo dose verification using using an amorphous silicon flat panel-type imager (a-Si EPIDs)
Introduction: Electronic portal imaging devices (EPIDs) could be used to dose verification of radiotherapy treatment plans. In vivo dose verification is performed to reduce differences found between dose delivered to the patient and the prescribed dose. The aim of this study was to perform a fast and efficient technique for the verification of delivered dose to the patient usin...
متن کاملEvaluation of the Influence of Exposure Index on Image Quality and Radiation Dose
Introduction: The introduction of digital radiography has led to a significant problem in terms of dose creep. To address this problem, manufacturers have established a set of exposure indicators (EI) as a feedback mechanism to safeguard against overexposure. The EI is the measure of incident exposure to the detector that is directly proportional to the signal-to-noise...
متن کاملA Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography
Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedures to reduce radiation dose without compromising the image quality. Material and Methods: This ex...