Newly Shaped Intra‐Aortic Balloons Improve the Performance of Counterpulsation at the Semirecumbent Position: An In Vitro Study

نویسندگان

  • Christina Kolyva
  • John R. Pepper
  • Ashraf W. Khir
چکیده

The major hemodynamic benefits of intra-aortic balloon pump (IABP) counterpulsation are augmentation in diastolic aortic pressure (Paug ) during inflation, and decrease in end-diastolic aortic pressure (ΔedP) during deflation. When the patient is nursed in the semirecumbent position these benefits are diminished. Attempts to change the shape of the IAB in order to limit or prevent this deterioration have been scarce. The aim of the present study was to investigate the hemodynamic performance of six new IAB shapes, and compare it to that of a traditional cylindrical IAB. A mock circulation system, featuring an artificial left ventricle and an aortic model with 11 branches and physiological resistance and compliance, was used to test one cylindrical and six newly shaped IABs at angles 0, 10, 20, 30, and 40°. Pressure was measured continuously at the aortic root during 1:1 and 1:4 IABP support. Shape 2 was found to consistently achieve, in terms of absolute magnitude, larger ΔedP at angles than the cylindrical IAB. Although ΔedP was gradually diminished with angle, it did so to a lesser degree than the cylindrical IAB; this diminishment was only 53% (with frequency 1:1) and 40% (with frequency 1:4) of that of the cylindrical IAB, when angle increased from 0 to 40°. During inflation Shape 1 displayed a more stable behavior with increasing angle compared to the cylindrical IAB; with an increase in angle from 0 to 40°, diastolic aortic pressure augmentation dropped only by 45% (with frequency 1:1) and by 33% (with frequency 1:4) of the drop reached with the cylindrical IAB. After compensating for differences in nominal IAB volume, Shape 1 generally achieved higher Paug over most angles. Newly shaped IABs could allow for IABP therapy to become more efficient for patients nursed at the semirecumbent position. The findings promote the idea of personalized rather than generalized patient therapy for the achievement of higher IABP therapeutic efficiency, with a choice of IAB shape that prioritizes the recovery of those hemodynamic indices that are more in need of support in the unassisted circulation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intra-aortic balloon pumping: does posture matter?

BACKGROUND The effect of the posture of semirecumbent patients on the hemodynamics and performance of intra-aortic balloon (IAB) were studied in vivo and in vitro. METHODS An IAB was inserted into a glass tube filled with saline, fully inflated and deflated using an intra-aortic balloon pump (IABP). Three successive cycles were filmed at 125 frames/s with the tube positioned at various angles...

متن کامل

Intra-aortic balloon counterpulsation in acute myocardial infarction: too few or too many?

The concept of counterpulsation was introduced over four decades ago by Clauss et al. (1) and adapted to intra-aortic balloon pumping (IABP) by Moulopaulus et al. (2) in 1962, and it remains the most common mode of cardiac assist. Intra-aortic balloon counterpulsation (IABCP) was first introduced into clinical practice by Kantrowitz et al. (3) in 1968 for the treatment of cardiogenic shock. Ini...

متن کامل

Effects of intra-aortic balloon counterpulsation on cardiac performance, oxygen consumption, and coronary blood flow in dogs.

The effect of intra-aortic counterpulsation (IACP) with a balloon upon myocardial oxygen consumption (MVo2), coronaiy blood flow (TCF), and left ventricular performance was studied in 23 anesthetized canine right heart bypass preparations at constant heart rate and cardiac output. In nonhypotensive, nonTCF-limited preparations, IACP produced a fall in left ventricular peak systolic pressure (LV...

متن کامل

Percutaneous aortic valve replacement. Part 3: Balloon counterpulsation of a novel temporary aortic valve.

A previously published two-part study described an engineering design of a percutaneous aortic valve (PAV) replacement system, which utilizes a novel temporary aortic valve (TAV) support to improve procedural outcomes and safety. Conceptually, this investigational approach can promote accurate PAV placement, procedural hemodynamic stability, smaller catheter delivery system, reduction in PAV re...

متن کامل

Staphylococcus aureus bacteremia associated with intra-aortic balloon pumps.

To the Editor: Since its first clinical use in the late 1960s, intra-aortic balloon pump counterpulsation has been a significant means of supporting an inadequate myocardium.' The use of an intra-aortic balloon pump has increased greatly preceding open heart surgery for cardiogenic shock and left ventricular failure. Additional indications include other conditions such as unstable angina refrac...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 40  شماره 

صفحات  -

تاریخ انتشار 2016