چکیده

INTRODUCTION During radiofrequency ablation to encircle or isolate the pulmonary veins (PVs), applications of radiofrequency energy within a PV may result in stenosis. The aim of this study was to determine whether monitoring of real-time impedance facilitates detection of inadvertent catheter movement into a PV. METHODS AND RESULTS In 30 consecutive patients (mean age 53 +/- 11 years) who underwent a left atrial ablation procedure, the three-dimensional geometry of the left atrium, the PVs, and their ostia were reconstructed using an electroanatomic mapping system. The PV ostia were identified based on venography, changes in electrogram morphology, and manual and fluoroscopic feedback as the catheter was withdrawn from the PV into the left atrium. Real-time impedance was measured at the ostium, inside the PV at approximately 1 and 3 cm from the ostium, in the left atrial appendage, and at the posterior left atrial wall. There was an impedance gradient from the distal PV (127 +/- 30 Omega) to the proximal PV (108 +/- 15 Omega) to the ostium (98 +/- 11 Omega) in each PV (P < 0.01). There was no significant impedance difference between the ostial and left atrial sites. During applications of radiofrequency energy, movement of the ablation catheter into a PV was accurately detected in 80% of the cases (20) when there was an abrupt increase of >/=4 Omega in real-time impedance. CONCLUSION There is a significant impedance gradient from the distal PV to the left atrium. Continuous monitoring of the real-time impedance facilitates detection of inadvertent catheter movement into a PV during applications of radiofrequency energy.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Catheter cryoablation of cardiac arrhythmias.

PURPOSE OF REVIEW The purpose of the review is to provide an update on the safety and efficacy of catheter cryoablation. RECENT FINDINGS Catheter cryoablation is a safe and clinically effective method for ablation of atrioventricular nodal reentrant supraventricular tachycardia. Although the acute procedural success rate of catheter cryoablation for this arrhythmia may be slightly lower than ...

متن کامل

Pulmonary Embolism as a Complication after radiofrequency catheter ablation Study: A Case Report

Pulmonary thromboembolism (PTE) is a fatal condition that may rarely occur due to complications of coronary catheter insertion. In this case report, a 41-year-old male was presented 48 hours after radiofrequency catheter ablation(RFCA) for the management of Wolf-Parkinson-White syndromewith acute onset of dyspnea, hemoptysis, and chest pain. The physical examination revealed coarse crackles in ...

متن کامل

Real-time 3-dimensional transesophageal echocardiography during left atrial radiofrequency catheter ablation for atrial fibrillation.

Left atrial radiofrequency catheter ablation has been recognized as an important treatment option for drugrefractory symptomatic atrial fibrillation.1 Recent consensus on technique favors catheter ablation directed to the left atrium near the pulmonary vein (PV) ostium to achieve PV isolation.2 However, ablation in the region of the ligament of Marshall (LOM) to achieve electrical isolation of ...

متن کامل

Radiofrequency Catheter Ablation for Atrial Fibrillation Elicited “Jackhammer Esophagus”: A New Complication Due to Vagal Nerve Stimulation?

Radiofrequency catheter ablation (RFCA) is a potentially curative method for treatment of highly symptomatic and drug-refractory atrial fibrillation (AF). However, this technique can provoke esophageal and nerve lesion, due to thermal injury. To our knowledge, there have been no reported cases of a newly described motor disorder, the Jackhammer esophagus (JE) after RFCA, independently of GERD. ...

متن کامل

Multielectrode Pulmonary Vein Isolation Versus Single Tip Wide Area Catheter Ablation for Paroxysmal Atrial Fibrillation

Catheter ablation (CA) is an accepted treatment option for symptomatic atrial fibrillation (AF). Historically, CA has been performed with a point-by-point technique, using standard focal radiofrequency catheters and 3-dimensional (3D) navigation systems, with a procedural end point of electric pulmonary vein isolation (PVI). However, this CA approach requires advanced training and operator skil...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of cardiovascular electrophysiology

دوره 15 6  شماره 

صفحات  -

تاریخ انتشار 2004