Comparison of four radiofrequency ablation systems at two target volumes in an ex vivo bovine liver model.
نویسندگان
چکیده
PURPOSE We aimed to validate actually achieved macroscopic ablation volumes in relation to calculated target volumes using four different radiofrequency ablation (RFA) systems operated with default settings and protocols for 3 cm and 5 cm target volumes in ex vivo bovine liver. MATERIALS AND METHODS Sixty-four cuboid liver specimens were ablated with four commercially available RFA systems (Radionics Cool-tip, AngioDynamic 1500X, Boston Scientific RF 3000, Celon CelonPower LAB): 16 specimens for each system; eight for 3 cm, and eight for 5 cm. Ablation diameters were measured, volumes were calculated, and RFA times were recorded. RESULTS For the 3 cm target ablation volume, all tested RFA systems exceeded the mathematically calculated volume of 14.14 cm3. For the 3 cm target ablation volume, mean ablation volume and mean ablation time for each RFA system were as follows: 28.5 ± 6.5 cm3, 12.0 ± 0.0 min for Radionics Cool-tip; 17.1 ± 4.9 cm3, 9.36 ± 0.63 min for AngioDynamic 1500X; 29.7 ± 11.7 cm3, 4.60 ± 0.50 min for Boston Scientific RF 3000; and 28.8 ± 7.0 cm3, 20.85 ± 0.86 min for Celon CelonPower LAB. For the 5 cm target ablation volume, Radionics Cool-tip (48.3 ± 9.9 cm3, 12.0 ± 0.0 min) and AngioDynamic 1500X (39.4 ± 16.2 cm3, 19.59 ± 1.13 min) did not reach the mathematically calculated target ablation volume (65.45 cm3), whereas Boston Scientific RF 3000 (71.8 ± 14.5 cm3, 9.15 ± 2.93 min) and Celon CelonPower LAB (93.9 ± 28.1 cm3, 40.21 ± 1.78 min) exceeded it. CONCLUSION While all systems reached the 3 cm target ablation volume, results were variable for the 5 cm target ablation volume. Only Boston Scientific RF 3000 and Celon CelonPower LAB created volumes above the target, whereas Radionics Cool-tip and AngioDynamic 1500X remained below the target volume. For the 3 cm target ablation volume, AngioDynamic 1500X with 21% deviation was closest to the target volume. For the 5 cm target volume Boston Scientific RF 3000 with 10% deviation was closest.
منابع مشابه
Moving-Shot versus Fixed Electrode Techniques for Radiofrequency Ablation: Comparison in an Ex-Vivo Bovine Liver Tissue Model
OBJECTIVE To compare the ablation characteristics of the moving-shot technique (MST) and the fixed electrode technique (FET) for radiofrequency (RF) ablation in an ex-vivo bovine liver tissue model. MATERIALS AND METHODS We performed RF ablation using FET in 110 bovine liver blocks using 11 different ablation times ranging from 5 seconds to 5 minutes (10 blocks per each time duration). Ten bo...
متن کاملAn Investigation of Brightness Changes in Ultrasound Images due to Temperature Variations in Liver Tissue during Radiofrequency Ablation
Introduction: one of the minimally invasive methods of treatmenting liver malignancies is Radiofrequency Ablation (RFA) which can be applied to primary or secondary tumors. Nevertheless, the disease recurrence is probable after RFA application partially due to the inadequate capability of temperature monitoring of the target tissue and visualizing the thermal damage. The goal of this work was t...
متن کاملNo-Touch Radiofrequency Ablation: A Comparison of Switching Bipolar and Switching Monopolar Ablation in Ex Vivo Bovine Liver
OBJECTIVE To evaluate the feasibility, efficiency, and safety of no-touch switching bipolar (SB) and switching monopolar (SM) radiofrequency ablation (RFA) using ex vivo bovine livers. MATERIALS AND METHODS A pork loin cube was inserted as a tumor mimicker in the bovine liver block; RFA was performed using the no-touch technique in the SM (group A1; 10 minutes, n = 10, group A2; 15 minutes, n...
متن کاملMultipolar radiofrequency ablation with internally cooled electrodes: experimental study in ex vivo bovine liver with mathematic modeling.
PURPOSE To evaluate the size and geometry of thermally induced coagulation by using multipolar radiofrequency (RF) ablation and to determine a mathematic model to predict coagulation volume. MATERIALS AND METHODS Multipolar RF ablations (n = 80) were performed in ex vivo bovine livers by using three internally cooled bipolar applicators with two electrodes on the same shaft. Applicators were ...
متن کاملLarge-volume radiofrequency ablation of ex vivo bovine liver with multiple cooled cluster electrodes.
Three methods of creating large thermal lesions with cool-tip cluster electrodes were compared. Three cluster electrodes were arranged 4 cm apart in a triangular array. Eight lesions were created ex vivo in fresh bovine liver (from a butcher) with each method: sequential ablation (three electrodes, 12 minutes each); simultaneous activation of electrodes (12 minutes); and rapid switching of powe...