Signal-to-noise ratio and parallel imaging performance of a 16-channel receive-only brain coil array at 3.0 Tesla.

نویسندگان

  • Jacco A de Zwart
  • Patrick J Ledden
  • Peter van Gelderen
  • Jerzy Bodurka
  • Renxin Chu
  • Jeff H Duyn
چکیده

The performance of a 16-channel receive-only RF coil for brain imaging at 3.0 Tesla was investigated using a custom-built 16-channel receiver. Both the image signal-to-noise ratio (SNR) and the noise amplification (g-factor) in sensitivity-encoding (SENSE) parallel imaging applications were quantitatively evaluated. Furthermore, the performance was compared with that of hypothetical coils with one, two, four, and eight elements (n) by combining channels in software during image reconstruction. As expected, both the g-factor and SNR improved substantially with n. Compared to an equivalent (simulated) single-element coil, the 16-channel coil showed a 1.87-fold average increase in brain SNR. This was mainly due to an increase in SNR in the peripheral brain (an up to threefold SNR increase), whereas the SNR increase in the center of the brain was 4%. The incremental SNR gains became relatively small at large n, with a 9% gain observed when n was increased from 8 to 16. Compared to the (larger) product birdcage head coil, SNR increased by close to a factor of 2 in the center, and by up to a factor of 6 in the periphery of the brain. For low SENSE acceleration (rate-2), g-factors leveled off for n>4, and improved only slightly (1.4% averaged over brain) going from n=8 to n=16. Improvements in g for n>8 were larger for higher acceleration rates, with the improvement for rate-3 averaging 12.0%.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Human Brain Imaging at 9.4 Tesla Using a Combination of Traveling Wave Excitation with a 15-channel Receive-only Array

Introduction The combination of a circularly polarized (CP) volume coil for spin excitation with a multi-channel receive-only array is a successful setup for routine human brain imaging at 7 Tesla. For reception, the use of multiple surface coils is important for a high SNR and for the utilization of parallel imaging techniques. For transmission, the main strength of the CP coil is the reasonab...

متن کامل

96-Channel receive-only head coil for 3 Tesla: design optimization and evaluation.

The benefits and challenges of highly parallel array coils for head imaging were investigated through the development of a 3T receive-only phased-array head coil with 96 receive elements constructed on a close-fitting helmet-shaped former. We evaluated several designs for the coil elements and matching circuitry, with particular attention to sources of signal-to-noise ratio (SNR) loss, includin...

متن کامل

Optimization of a high sensitivity MRI receive coil for parallel human brain imaging

Two eight-channel MRI receive-only coils were developed to provide whole-brain coverage at 1.5 T and 3.0 T field strength, respectively. Objectives were an image signal-to-noise ratio superior to standard designs throughout the human brain, as well as high parallel imaging performance. Electro-magnetic field simulations were used to determine array diameter and inter-element coil gap. Low mutua...

متن کامل

A geometrically adjustable 16-channel transmit/receive transmission line array for improved RF efficiency and parallel imaging performance at 7 Tesla.

A novel geometrically adjustable transceiver array system is presented. A key feature of the geometrically adjustable array was the introduction of decoupling capacitors that allow for automatic change in capacitance dependent on neighboring resonant element distance. The 16-element head array version of such an adjustable coil based on transmission line technology was compared to fixed geometr...

متن کامل

A Specialized Multi-Transmit Head Coil for High Resolution fMRI of the Human Visual Cortex at 7T

PURPOSE To design, construct and validate radiofrequency (RF) transmit and receive phased array coils for high-resolution visual cortex imaging at 7 Tesla. METHODS A 4 channel transmit and 16 channel receive array was constructed on a conformal polycarbonate former. Transmit field efficiency and homogeneity were simulated and validated, along with the Specific Absorption Rate, using [Formula:...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Magnetic resonance in medicine

دوره 51 1  شماره 

صفحات  -

تاریخ انتشار 2004