Applanation and Schi0tz tonometer standardizations for the owl monkey eye with a new technique for measuring episcleral venous pressure

نویسنده

  • James L. Mims
چکیده

A Schi0tz Pi calibration by open manometer was performed on 18 owl monkey eyes cannulated in situ. Pi values for the owl monkey corresponded closely to those of the 1955 Friedenwald nomogram. The average ocular rigidity measured in 16 additional eyes with the use of clinical techniques was 0.0173 (± 0.0015). A manometric calibration for a 4 mm. diameter applanation face was done similarly on 21 eyes. The following standardization formula ivas generated: Pressure in mm. Hg = 0.641 (tonometric scale reading) + 2.56 mm. Hg. An attempt ivas made to compare the two tonometric pressure standardizations. In 16 intact eyes, Schi0tz PO pressures, corrected for ocular rigidity, averaged 0.77 ± 0.47 mm. Hg higher than corresponding applanation pressures. An improved method for measuring episcleral venous pressure, xoith the use of a new synthetic polyurethane pressure chamber membrane, is also reported.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Applanation and Schiotz tonometer standardizations for the owl monkey eye with a new technique for measuring episcleral venous pressure.

A Schi0tz Pi calibration by open manometer was performed on 18 owl monkey eyes cannulated in situ. Pi values for the owl monkey corresponded closely to those of the 1955 Friedenwald nomogram. The average ocular rigidity measured in 16 additional eyes with the use of clinical techniques was 0.0173 (± 0.0015). A manometric calibration for a 4 mm. diameter applanation face was done similarly on 21...

متن کامل

Re-evaluation of the Schiotz tonometer calibration.

The calibration scale for the Schi0tz tonometers toas tested by comparing Goldmann applanation and Schi0tz measurements on the eyes of 906 patients. The study differed from several previous studies in that both the applanation and Schi0tz measurements were made with the subjects lying on their backs, thus eliminating errors due to change in intraocular pressure as the subject changed from one p...

متن کامل

Pressure measurement: which tonometer?

In the search for more accurate ways to determine intraocular pressure, the Halberg applanation tonometer may represent an alternative to the Schi0tz tonometer for routine screening pressure measurements. Although between 100 to 150 measurements are generally required to gain proficiency with the Halberg, the instrument is accurate in patients with normal corneas even with very abnormal scleral...

متن کامل

EPISCLERAL VENOUS PRESSURE IN TONOGRAPHY * t BY ARTHUR

DURING the 11 years in which the clinical measurement of facility of outflow by means of tonography has been practised, various workers have sought to elucidate and correct the sources of error inherent in this method. One of the assumptions originally made concerning tonography was that the episcleral venous pressure was not disturbed during the measurement. Friedenwald (Grant, 1951) and Becke...

متن کامل

The effect of miotics on the intraocular pressure of conscious owl monkeys.

The intraocular pressure of conscious, unsedated owl monkeys (Aotus trivirgatus) was measured with an applanation tonometer. Untreated eyes of the conscious animals were found to have higher values than those reported for owl monkeys anesthetized with pentobarbitone. Locally applied pilocarpine, carbachol, and oxotremorine gave concentration-related reduction in pressure, oxotremorine being the...

متن کامل