Implementation of Constant Dose Rate and Constant Angular Spacing Intensity-modulated Arc Therapy for Cervical Cancer by Using a Conventional Linear Accelerator
نویسندگان
چکیده
BACKGROUND Volumetric-modulated arc therapy (VMAT) can only be implemented on the new generation linacs such as the Varian Trilogy® and Elekta Synergy®. This prevents most existing linacs from delivering VMAT. The purpose of this study was to investigate the feasibility of using a conventional linear accelerator delivering constant dose rate and constant angular spacing intensity-modulated arc therapy (CDR-CAS-IMAT) for treating cervical cancer. METHODS Twenty patients with cervical cancer previously treated with intensity-modulated radiation therapy (IMRT) using Varian Clinical 23EX were retreated using CDR-CAS-IMAT. The planning target volume (PTV) was set as 50.4 Gy in 28 fractions. Plans were evaluated based on the ability to meet the dose volume histogram. The homogeneity index (HI), target volume conformity index (CI), the dose to organs at risk, radiation delivery time, and monitor units (MUs) were also compared. The paired t-test was used to analyze the two data sets. All statistical analyses were performed using SPSS 19.0 software. RESULTS Compared to the IMRT group, the CDR-CAS-IMAT group showed better PTV CI (0.85 ± 0.03 vs. 0.81 ± 0.03, P = 0.001), clinical target volume CI (0.46 ± 0.05 vs. 0.43 ± 0.05, P = 0.001), HI (0.09 ± 0.02 vs. 0.11 ± 0.02, P = 0.005) and D95 (5196.33 ± 28.24 cGy vs. 5162.63 ± 31.12 cGy, P = 0.000), and cord D2 (3743.8 ± 118.7 cGy vs. 3806.2 ± 98.7 cGy, P = 0.017) and rectum V40 (41.9 ± 6.1% vs. 44.2 ± 4.8%, P = 0.026). Treatment time (422.7 ± 46.7 s vs. 84.6 ± 7.8 s, P = 0.000) and the total plan Mus (927.4 ± 79.1 vs. 787.5 ± 78.5, P = 0.000) decreased by a factor of 0.8 and 0.15, respectively. The IMRT group plans were superior to the CDR-CAS-IMAT group plans considering decreasing bladder V50 (17.4 ± 4.5% vs. 16.6 ± 4.2%, P = 0.049), bowel V30 (39.6 ± 6.5% vs. 36.6 ± 7.5%, P = 0.008), and low-dose irradiation volume; there were no significant differences in other statistical indexes. CONCLUSIONS Patients with cervical cancer treated with CDR-CAS-IMAT using Varian Clinical 23EX can get equivalent or superior dose distribution compared to those treated with IMRT. CDR-CAS-IMAT has a less treatment time and MU, which can reduce the uncertainty factor and patient discomfort in treatment.
منابع مشابه
Variable dose rate single-arc IMAT delivered with a constant dose rate and variable angular spacing.
Single-arc intensity-modulated arc therapy (IMAT) has gained worldwide interest in both research and clinical implementation due to its superior plan quality and delivery efficiency. Single-arc IMAT techniques such as the Varian RapidArc deliver conformal dose distributions to the target in one single gantry rotation, resulting in a delivery time in the order of 2 min. The segments in these tec...
متن کاملDesign, Simulation and Manufacture of a Multileaf Collimator to Confirm the Target Volumes in Intensity Modulated Radiation Therapy
Introduction: Intensity modulated radiation therapy (IMRT) is one of the cancer treatment methods. It is important to selectively aim at the target in this way, which can be performed using a multileaf collimator (MLC). Materials and Methods: In order to specifically irradiate the target volume in radiotherapy to reduce the patient absorbed dose, the use of multileaf collimator has been investi...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملDose Distribution Analysis of Rapid Arc and Intensity Modulated Radiotherapy Plan in Head and Neck Cancer
Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nom...
متن کاملOptimization of prostate cancer radiotherapy using of a spacer gel, volumetric modulated arc therapy and a single biological organ at risk objective
Background: The aim was to evaluate the benefit of technical advances for treatment planning: introduction of a hydrogel spacer, VMAT (volumetric modulated arc therapy) and a single biological organ at risk objective for the rectum and bladder. Initial standard was a step-and-shoot IMRT (intensity modulated radiotherapy) without a spacer and conventional organ at risk objectives. Material...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 129 شماره
صفحات -