In-vivo evaluation of the kinematic behavior of an artificial medial meniscus implant: A pilot study using open-MRI.

نویسندگان

  • Tineke De Coninck
  • Jonathan J Elsner
  • Eran Linder-Ganz
  • Michiel Cromheecke
  • Maoz Shemesh
  • Wouter Huysse
  • René Verdonk
  • Koenraad Verstraete
  • Peter Verdonk
چکیده

BACKGROUND In this pilot study we wanted to evaluate the kinematics of a knee implanted with an artificial polycarbonate-urethane meniscus device, designed for medial meniscus replacement. The static kinematic behavior of the implant was compared to the natural medial meniscus of the non-operated knee. A second goal was to evaluate the motion pattern, the radial displacement and the deformation of the meniscal implant. METHODS Three patients with a polycarbonate-urethane implant were included in this prospective study. An open-MRI was used to track the location of the implant during static weight-bearing conditions, within a range of motion of 0° to 120° knee flexion. Knee kinematics were evaluated by measuring the tibiofemoral contact points and femoral roll-back. Meniscus measurements (both natural and artificial) included anterior-posterior meniscal movement, radial displacement, and meniscal height. FINDINGS No difference (P>0.05) was demonstrated in femoral roll-back and tibiofemoral contact points during knee flexion between the implanted and the non-operated knees. Meniscal measurements showed no significant difference in radial displacement and meniscal height (P>0.05) at all flexion angles, in both the implanted and non-operated knees. A significant difference (P ≤ 0.05) in anterior-posterior movement during flexion was observed between the two groups. INTERPRETATION In this pilot study, the artificial polycarbonate-urethane implant, indicated for medial meniscus replacement, had no influence on femoral roll-back and tibiofemoral contact points, thus suggesting that the joint maintains its static kinematic properties after implantation. Radial displacement and meniscal height were not different, but anterior-posterior movement was slightly different between the implant and the normal meniscus.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of MRI in Comparison with Arthroscopic Findings in the Diagnosis of Anterior Cruciate Ligament and Meniscus Tears

Background: Magnetic resonance imaging (MRI) is of great aid in the diagnosis of knee lesions. Nevertheless, arthroscopy has remained the reference standard for the diagnosis of internal derangements of the knee, against which alternative diagnostic modalities should be compared. This study was designed to assess the value of MRI and clinical examination in the diagnosis of Anterior Cruciate Li...

متن کامل

شیوع منیسکال‌فلونس و انواع تروماهای منیسکولیگامانی زانو در بیماران مراجعه‌کننده به مرکز MRI کوثر بیمارستان امام رضا مشهد

Background and Aim: Meniscal flounceis a rare, wavy shaped folding along the free edge of the meniscus that observed in .MRI images. This is a feature of transient distortion of a normal meniscus but it can be regarded as a pseudo- meniscal tear. The aimof the currentstudy was to evaluate the incidence of this normal variant. Materials and Methods: A total number of1001 knee MRI images. were pr...

متن کامل

ارزیابی حساسیت و ویژگی MRI در مقایسه با آرتروسکوپی در ضایعات زانو

Background: The knee is the most commonly injured joint. The internal components of the knee include the meniscuses, the anterior cruciate ligament (ACL) and posterior cruciate ligaments (PCL). Magnetic resonance imaging (MRI) is now used widely because of its high accuracy and noninvasiveness. Our objective was to diagnose internal knee component injuries by MRI and compare its diagnostic valu...

متن کامل

Kinematics of Normal Menisci during Knee Flexion

The human knee menisci are important structures to, proper knee joint function. At present, little data exists describing the meniscal kinematics of the intact knee. Accordingly, the purpose of this study was to determine in-vivo meniscal kinematic motion of the non-injured knee during unloaded knee flexion through use of magnetic resonance imaging (MRI). Seven females and three males (mean age...

متن کامل

Loop-Post Construct, A Novel Technique for Medial Meniscus Root Repair

Surgical reattachment of medial meniscus posterior root tear (MMPRT) with transtibial sutures can delay the presenceof medial knee joint compartment osteoarthritis. Most suture configurations are placed five mm away from the tornmargin in the meniscal substance which is already degenerated and may decrease the pull out strengths of repairconstruct. The number of meniscus penet...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical biomechanics

دوره 29 8  شماره 

صفحات  -

تاریخ انتشار 2014