Specific absorption rate in neonates undergoing magnetic resonance procedures at 1.5 T and 3 T

نویسندگان

  • Shaihan J. Malik
  • Arian Beqiri
  • Anthony N. Price
  • Jose Nuno Teixeira
  • Jeffrey W. Hand
  • Joseph V. Hajnal
چکیده

MRI is finding increased clinical use in neonatal populations; the extent to which electromagnetic models used for quantification of specific absorption rate (SAR) by commercial MRI scanners accurately reflect this alternative scenario is unclear. This study investigates how SAR predictions relating to adults can be related to neonates under differing conditions when imaged using 1.5 T and 3 T MRI scanners. Electromagnetic simulations were produced in neonatal subjects of different sizes and positions within a generic MRI body transmit device operating at both 64 MHz and 128 MHz, corresponding to 1.5 T and 3 T MRI scanners, respectively. An adult model was also simulated, as was a spherical salt-water phantom, which was also used in a calorimetry experiment. The SAR in neonatal subjects was found to be less than that experienced in an adult in all scenarios; however, the overestimation factor was variable. For example a 3 T body scan resulting in local 10 g SAR of 10.1 W kg(-1) in an adult would deposit 2.6 W kg(-1) in a neonate: an approximately fourfold difference. The SAR experienced by neonatal subjects undergoing MRI is lower than that in adults in equivalent situations. If the safety of such procedures is assessed using adult-appropriate models then the result is a conservative estimate.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Clinical Experience with 3.0 T MR for Cardiac Imaging in Patients: Comparison to 1.5 T using Individually Optimized Imaging Protocols

Cardiovascular magnetic resonance imaging (MRI) has gained widespread acceptance for the assessment of cardiovascular disease (1-4). Recent developments in 3.0 T MRI technology have led to an increased interest in high-field cardiac imaging. Because the signal-to-noise (SNR) scales approximately linearly with the magnetic field strength, B0, an improvement in the SNR can be expected at 3.0 T in...

متن کامل

Assessment of MRI issues for the Argus II retinal prosthesis.

OBJECTIVE The objective was to evaluate magnetic resonance imaging (MRI) issues (magnetic field interactions, heating, artifacts and functional alterations) at 1.5 T and 3 T for the Argus II Retinal Prosthesis (Second Sight Medical Products, Sylmar, CA, USA). MATERIALS AND METHODS Standardized protocols were used to assess magnetic field interactions (translational attraction and torque; 3 T,...

متن کامل

Prosthetic heart valves: evaluation of magnetic field interactions, heating, and artifacts at 1.5 T.

The purpose of this study was to use ex vivo testing techniques to determine the magnetic resonance imaging (MRI) safety aspects for 32 different heart valve prostheses that had not been evaluated previously in association with the 1.5-T MR environment. Ex vivo testing was performed using previously described techniques for the evaluation of magnetic field interactions (deflection angle and tor...

متن کامل

Prosthetic heart valves and annuloplasty rings: assessment of magnetic field interactions, heating, and artifacts at 1.5 Tesla.

The purpose of this study was to determine the magnetic resonance (MR) safety aspects and artifacts for three different heart valve prostheses and two different annuloplasty rings that have not been evaluated previously in association with the 1.5-T MR environment. Ex vivo testing was performed using previously described techniques for the evaluation of magnetic field interactions (deflection a...

متن کامل

Ultrafast magnetic resonance imaging of the neonate in a magnetic resonance-compatible incubator with a built-in coil.

BACKGROUND Magnetic resonance (MR) imaging of the neonate is important clinically, because this group of patients often has complex and multiple problems due to prematurity and developmental abnormalities. MR imaging usually involves moving neonates away from their controlled environment to the scanner. OBJECTIVE In this study we present the results of our initial experience with an MR-compat...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 28  شماره 

صفحات  -

تاریخ انتشار 2015