Triprolidine (Trip) ion selective electrodes of three types: the conventional polymer membrane (I), graphite coated electrode (II) and carbon paste electrode (III), have been prepared, based on the ion pair of triprolidine hydrochloride with sodium tetraphenylborate. The electrodes exhibit a linear response with a mean calibration graph slope of 56.12, 55.00 and 54.32 mV decade(-1) at 25 degrees C for I, II and III, respectively, within the concentration ranges 1.96 x 10(-5) - 1.00 x 10(-2) M for I and 3.84 x 10(-5) - 1.00 x 10(-2) M for II and III. The detection limits are 1.13+/-0.13 x 10(-5), 1.70+/-0.06 x 10(-5) and 1.78+/-0.05 x 10(-5) M for the three electrodes, respectively. The change of pH within the ranges 4.85 - 8.75 and 4.70 - 8.50 for I and III, respectively, did not affect the electrode performance. The standard electrode potentials were determined at different temperatures and were used to calculate the isothermal coefficient of the electrode. The electrodes showed a very good selectivity for Trip with respect to a large number of inorganic cations and compounds. The standard addition method was applied to the determination of TripCl in pure solution, pharmaceutical preparations, and urine samples.
منابع مشابه
Ion-Selective Carbon Paste Electrode Based on 2-Amino-N-benzthioazolyl Benzamide (ABTB) for Determination of Copper (II) by Potentiometric Method
2-Amino-N-benzthioazolyl benzamide (ABTB) was synthesized, characterized and used for thefabrication of a potentiometric sensor for Cu2+ metal ions. The electrode exhibits linear response to Cu(II) over a wide concentration range (4.79×10 -8 – 1.85×10 -1 M) with Nernstian slope of 30 ± 1.5 mVper decade. The electrode can be used in the pH range from 2 to 9. It has a fast response time of about1...
متن کاملIon Selective Carbon Paste Electrode For The Determination Of Flavoxate Muscle Relaxant Drug In Pharmaceutical Formulation
Abstract The utility of carbon paste electrode for the determination of flavoxate HCl modified with flavoxate-tetraphenylborate (FLX-TPB) and flavoxate-phosphotungestic acid (FLX-PTA) ion-pairs in batch mode is demonstrated. The electrodes revealed a Nernstian response over a wide concentration ranges 1.39×10-5-1x10-2 mol L-1 and1×10-5-1x10-2 mol L-1 using FLX-TPB and FLX-PTA, respectively. The...
متن کاملGreen Potentiometric Method for Determination of Triprolidine Hydrochloride and Pseudoephedrine Hydrochloride in their Different Pharmaceutical Matrices using Liquid and Solid Contact Gold Electrodes
Factors affecting the performance of polyvinyl chloride based ion selective electrode such as using different cationic exchangers and ionophores as well as sensor fabrication procedures were recently the scope of study by several researchers. This work presents an environment friendly electro-analytical method for the determination of Triprolidine HCl (TRI) and Pseudoephedrine HCl (PSE). A comp...
متن کاملFabrication of an Electrochemical Sensor Based on a New Nano-ion Imprinted Polymer for Highly Selective and Sensitive Determination of Molybdate
In this work a new chemically modified carbon paste electrode was constructed for accurate, simple, sensitive and selective determination of molybdenum (VI) ions. The results of modified electrode by an ion imprinted polymer were compared with those obtained with carbon paste electrode. The results showed the stripping peak currents had a dramatic increase at the modified electrode. Under the o...
متن کاملDevelopment of novel potentiometric sensors for determination of lidocaine hydrochloride in pharmaceutical preparations, serum and urine samples
This article is focused on the determination of lidocaine hydrochloride as a local anaesthetic drug. A potentiometric method based on modified screen-printed and modified carbon paste ion-selective electrodes was described for the determination of lidocaine hydrochloride in different pharmaceutical preparations. It was based on potentiometric titration of lidocaine hydrochloride using modified ...