A retrieval study of capital hip prostheses with titanium alloy femoral stems.

نویسندگان

  • L R Mcgrath
  • D L Shardlow
  • E Ingham
  • M Andrews
  • J Ivory
  • M H Stone
  • J Fisher
چکیده

We have examined 26 retrieved, failed titanium-alloy femoral stems. The clinical details, radiological appearances and the histology of the surrounding soft tissues in each patient were also investigated. The stems were predominantly of the flanged design and had a characteristic pattern of wear. A review of the radiographs showed a series of changes, progressive with time. The first was lateral debonding with subsidence of the stem. This was followed by calcar resorption and fragmentation or fracture of the cement. Finally, osteolysis was seen, starting with a radiolucency at the cement-bone interface and progressing to endosteal cavitation. Three histological appearances were noted: granulomatous, necrobiotic and necrotic. We suggest that an unknown factor, possibly related to the design of the stem, caused it to move early. After this, micromovement at the cement-stem interface led to the generation of particulate debris and fracture of the cement. A soft-tissue reaction to the debris resulted in osteolysis and failure of fixation of the prostheses.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Hydroxyapatite coating on carbon composite hip implants in dogs.

In 33 dogs we implanted femoral stems made either of carbon composite, some coated with hydroxyapatite, or of titanium alloy with a porous coating. Osseointegration was greater in the hydroxyapatite-coated than in the un-coated stems (p less than 0.001). Push-out tests, at an average of 7.2 months after implantation, showed a six-fold increase in interface shear strength and a twelve-fold incre...

متن کامل

Fretting and Corrosion Damage at the Head-Neck Taper is Reduced with Ceramic Femoral Heads: A Retrieval Study

Introduction: Modular head-neck total hip arthroplasty (THA) offers several benefits for orthopaedic surgeons, including intraoperative flexibility and the ability to change the head at revision surgery [1]. However, modularity at the head-neck taper, especially the use of metallic heads and stems, can lead to fretting and corrosion at the taper junction, leading to metal ions and debris releas...

متن کامل

Effects of material properties of femoral hip components on bone remodeling.

Bone loss around femoral hip stems is one of the problems threatening the long-term fixation of uncemented stems. Many believe that this phenomenon is caused by reduced stresses in the bone (stress shielding). In the present study the mechanical consequences of different femoral stem materials were investigated using adaptive bone remodeling theory in combination with the finite element method....

متن کامل

Failure of the uncoated titanium ProxiLock femoral hip prosthesis.

UNLABELLED New prostheses should be evaluated for stability and clinical performance. In a prospective randomized clinical trial, we implanted 22 titanium (Ti) and 20 hydroxyapatite-coated (HA) ProxiLock femoral hip prostheses during total hip arthroplasty in 42 patients. The patients were followed for 24 months with clinical, radiographic and radiostereometric analysis. Full weightbearing was ...

متن کامل

Wear particles and ions from cemented and uncemented titanium-based hip prostheses—A histological and chemical analysis of retrieval material

Wear debris-induced inflammation is considered to be the main cause for periprosthetic osteolysis in total hip replacements (THR). The objective of this retrieval study was to examine the tissue reactions and exposure to metal ions and wear particles in periprosthetic tissues and blood samples from patients with titanium (Ti)-based hip prostheses that were revised due to wear, osteolysis, and/o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 83 8  شماره 

صفحات  -

تاریخ انتشار 2001