Fresh Blood Imaging using variable TR and variable refocusing flip angle for non-contrast peripheral MR angiography at 3T: a 7-Minute iliac to calf station run-offs scan
نویسندگان
چکیده
Background Non-contrast Fresh Blood Imaging (FBI) scan time can be greatly reduced using variable TR for imaging the peripheral artery without contrast infusion, in which each slice encoding (SE) and the followed echo train are played in a variable TR (vTR) (TR=n*RR) pattern. In vTR FBI, the lower limit of TR depends on the length and timing of data acquisition, and it is also limited by SAR, which is mostly affected by refocusing flip angle, echo train length, and echo spacing. Often in the case of short TR FSE acquisition, SAR can be an issue for the patient with fast heart rate. To reduce SAR so that the shortest TR can be enabled for the slice encodings at the kz edge, variable refocusing flip angle (vFA) is proposed in this work. The optimized vTR FBI sequence with vFA was applied on volunteers to achieve a 7-minutes 3 station run-off scan.
منابع مشابه
ECG gated 3D single shot FSE with variable TR for non-contrast peripheral MRA at 3T
Background ECG gated 3D single shot fast spin echo (SSFSE) sequence can be used to acquire systolic and diastolic data and the subtracted image gives the arterial blood signal. This technique, also called Fresh Blood Imaging (FBI), has been applied to image the peripheral artery without contrast infusion, in which each slice encoding (SE) step is acquired in a fixed TR (TR=n*RR). Ideally, TR sh...
متن کاملComparison of nonenhanced MR angiographic subtraction techniques for infragenual arteries at 1.5 T: a preliminary study.
PURPOSE To evaluate diagnostic performance of three nonenhanced methods: variable-refocusing-flip angle (FA) fast spin-echo (SE)-based magnetic resonance (MR) angiography (variable FA MR) and constant-refocusing-FA fast SE-based MR angiography (constant-FA MR) and flow-sensitive dephasing (FSD)-prepared steady-state free precession MR angiography (FSD MR) for calf arteries, with dual-injection ...
متن کاملReduction in Artifacts in the Liver at 3T Using a 3D T2-Weighted Variable Flip-Angle Sequence (SPACE)
Introduction: 3T MRI systems provide an inherent increase in SNR that can be used for gains in spatial and/or temporal resolution. However, a limitation of imaging at 3T is an increase in specific absorption rate (SAR), particularly in sequences with large flip angles and many refocusing pulses, as occurs with many 180o refocusing pulses of a turbo spin echo sequence. A recently developed seque...
متن کاملAssessment of Non-Contrast Angiography in Diabetic Patients
Purpose Studies have shown that diabetic patients are four times more likely to develop peripheral artery disease (PAD) [1]. In addition, diabetes is the number one cause of kidney failure [2]. The combination of PAD onset and high risk of kidney failure associated with diabetes necessitates frequent angiographic exams for these patients. Contrast-enhanced MR Angiography (CE-MRA) has been shown...
متن کاملFlow sensitivity analysis of variable refocusing angle 3D FSE
Introduction: FSE sequences can be adapted either to maintain brightness or to suppress the signal of moving blood. Subtraction methods with FSE can generate angiographic images without injection of contrast agents [1]. 3D FSE variants with variable refocusing angle (VRA) of the echo train are naturally suitable for high scanning efficiency and SAR reduction [2]. We analyze sequence parameter m...