Gold-tip electrodes--a new "deep lesion" technology for catheter ablation? In vitro comparison of a gold alloy versus platinum-iridium tip electrode ablation catheter.

نویسندگان

  • Thorsten Lewalter
  • Alexander Bitzen
  • Sabine Wurtz
  • Robert Blum
  • Karsten Schlodder
  • Alexander Yang
  • Lars Lickfett
  • Jörg O Schwab
  • Jan W Schrickel
  • Klaus Tiemann
  • Markus Linhart
  • Endre Zima
  • Bela Merkely
  • Berndt Lüderitz
چکیده

Gold-tip electrodes. Radiofrequency (RF) catheter ablation is widely used to induce focal myocardial necrosis using the effect of resistive heating through high-frequency current delivery. It is current standard to limit the target tissue-electrode interface temperature to a maximum of 60-70 degrees C to avoid char formation. Gold (Au) exhibits a thermal conductivity of nearly four times greater than platinum (Pt-Ir) (3.17 W/cm Kelvin vs 0.716 W/cm Kelvin), it was therefore hypothesized that RF ablation using a gold electrode would create broader and deeper lesions as a result of a better heat conduction from the tissue-electrode interface and additional cooling of the gold electrode by "heat loss" to the intracardiac blood. Both mechanisms would allow applying more RF power to the tissue before the electrode-tissue interface temperature limit is reached. To test this hypothesis, we performed in vitro isolated liver and pig heart investigations comparing lesion depths of a new Au-alloy-tip electrode to standard Pt-Ir electrode material. Mean lesion depth in liver tissue for Pt-Ir was 4.33+/-0.45 mm (n=60) whereas Au electrode was able to achieve significantly deeper lesions (5.86+/-0.37 mm [n=60; P<0.001]). The mean power delivered using Pt-Ir was 6.95+/-2.41 W whereas Au tip electrode delivered 9.64+/-3.78 W indicating a statistically significant difference (P<0.05). In vitro pig heart tissue Au ablation (n=20) increased significantly the lesion depth (Au: 4.85+/-1.01 mm, Pt-Ir: 2.96+/-0.81 mm, n=20; P<0.001). Au tip electrode again applied significantly more power (P<0.001). Gold-tip electrode catheters were able to induce deeper lesions using RF ablation in vitro as compared to Pt-Ir tip electrode material. In liver and in pig heart tissue, the increase in lesion depth was associated with a significant increase in the average power applied with the gold electrode at the same level of electrode-tissue temperature as compared to platinum material.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Is 8-mm more effective than 4-mm tip electrode catheter for ablation of typical atrial flutter?

BACKGROUND The prospective, randomized study comparing 4- with 8-mm tip electrodes for radiofrequency linear ablation of typical atrial flutter is not available. METHODS AND RESULTS A total of 104 consecutive patients with typical atrial flutter were randomly assigned to undergo radiofrequency linear ablation using a 4- (Group I, n=54) or 8-mm tip electrode (Group II, n=50) catheter (temperat...

متن کامل

Lesion dimensions during temperature-controlled radiofrequency catheter ablation of left ventricular porcine myocardium: impact of ablation site, electrode size, and convective cooling.

BACKGROUND It is important to increase lesion size to improve the success rate for radiofrequency ablation of ischemic ventricular tachycardia. This study of radiofrequency ablation, with adjustment of power to approach a preset target temperature, ie, temperature-controlled ablation, explores the effect of catheter-tip length, ablation site, and convective cooling on lesion dimensions. METHO...

متن کامل

Low rate of asymptomatic cerebral embolism and improved procedural efficiency with the novel pulmonary vein ablation catheter GOLD: results of the PRECISION GOLD trial

AIMS This prospective, multicentre study (PRECISION GOLD) evaluated the incidence of asymptomatic cerebral embolism (ACE) after pulmonary vein isolation (PVI) using a new gold multi-electrode radiofrequency (RF) ablation catheter, pulmonary vein ablation catheter (PVAC) GOLD. Also, procedural efficiency of PVAC GOLD was compared with ERACE. The ERACE study demonstrated that a low incidence of A...

متن کامل

AF ablation: Single shot multielectrode or multishot single electrode?

There is no doubt that catheter ablation is established as an important tool in the armoury used in the management of symptomatic, drug-resistant atrial fibrillation [1]. Effective pulmonary vein isolation continues to be the main foundation of this approach [2]. Complete isolation of the pulmonary veins is associated with fewer atrial fibrillation recurrences when compared with incomplete lesi...

متن کامل

Temperature-guided radiofrequency catheter ablation with very large distal electrodes.

BACKGROUND Previous studies have shown that the size of lesions produced by radiofrequency catheter ablation correlates with the temperature and surface area of the electrode-tissue interface. The purpose of the present study was to compare the effects of ablation using very large distal electrodes (8F, 8 and 12 mm long) with those made by a conventional radiofrequency ablation catheter (distal...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of cardiovascular electrophysiology

دوره 16 12  شماره 

صفحات  -

تاریخ انتشار 2005