Calibration of Electret-Based Integral Radon Monitors Using NIST Polyethylene-Encapsulated 226Ra/222Rn Emanation (PERE) Standards
نویسندگان
چکیده
The recently developed 222Rn emanation standards that are based on polyethylene-encapsulated 226Ra solutions were employed for a first field-measurement application test to demonstrate their efficacy in calibrating passive integral radon monitors. The performance of the capsules was evaluated with respect to the calibration needs of electret ionization chambers (E-PERM®, Rad Elec Inc.). The encapsulated standards emanate well-characterized and known quantities of 222Rn, and were used in two different-sized, relatively-small, accumulation vessels (about 3.6 L and 10 L) which also contained the deployed electret monitors under test. Calculated integral 222Rn activities from the capsules over various accumulation times were compared to the averaged electret responses. Evaluations were made with four encapsulated standards ranging in 226Ra activity from approximately 15 Bq to 540 Bq (with 222Rn emanation fractions of 0.888); over accumulation times from 1 d to 33 d; and with four different types of E-PERM detectors that were independently calibrated. The ratio of the electret chamber response ERn to the integral 222Rn activity IRn was constant (within statistical variations) over the variables of the specific capsule used, the accumulation volume, accumulation time, and detector type. The results clearly demonstrated the practicality and suitability of the encapsulated standards for providing a simple and readily-available calibration for those measurement applications. However, the mean ratio ERn/IRn was approximately 0.91, suggesting a possible systematic bias in the extant E-PERM calibrations. This 9 % systematic difference was verified by an independent test of the E-PERM calibration based on measurements with the NIST radon-in-water standard generator.
منابع مشابه
Radon monitor calibration using NIST radon emanation standards: steady flow method.
The National Institute of Standards and Technology (NIST) polyethylene-encapsulated 226Ra/222Rn emanation (PERE) standards (old SRM 4968 and new SRMs 4971, 4972, and 4973) provide precise radon emanation rate, certified to a high degree of accuracy (approximately to 2%). Two new SRM 4973 standards containing totally 1036 Bq (0.028 microCi) of 226Ra, emanate 0.114 Bq (3.08 pCi) of 222Rn per min....
متن کاملAPPLICATION OF NIST ^Rn EMANATION STANDARDS FOR CALIBRATING ^Rn MONITORS
Several years of research in developing a radon emanation standard by the National Institute of Standards and Technology (NIST) has led to NIST Radon Emanation Standard SRM 4968 (Journal of Research of the MST 100:629-639; 1996). This new standard is based on a polyethylene -encapsulated ""Ra solution that has been demonstrated to emanate a well-characterized and known quantity "̂Rn when employe...
متن کاملThe NIST Primary Radon-222 Measurement System
Within the United States, the national standard for radon measurements is embodied in a primary radon measurement system that has been maintained for over 50 years to accurately measure radon (222Rn) against international and national radium (226Ra) standards. In turn, all of the radon measurements made at the National Institute of Standards and Technology (NIST) and the radon transfer calibrat...
متن کاملNist Traceable Radon Calibration System for Calibrating
The NIST (National Institute of Standards and Technology) has recently made available 222Rn emanation standards for test and evaluation. The NIST certified parameters include the ^ ~ a strength and the emanation coefficient. When such a source is loaded into a radon leak tight enclosure of a known volume, 222~n accumulates over time and it is possible to calculate precisely the time integrated ...
متن کاملRadon diffusion and the emanation fraction for NIST polyethylene capsules containing radium solution
The national standard for radon (Rn) measurements in the United States is based on Standard Reference Materials (SRMs) prepared at the National Institute of Standards and Technology (NIST), Gaithersburg, MD. NIST radon standards are hermetically sealed polyethylene capsules, filled with radium (Ra) solution. In preparation of standard reference materials for Rn, it is necessary to understand pr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 100 شماره
صفحات -
تاریخ انتشار 1995