Clinical assessment of a continuous intraarterial blood gas monitoring system.

نویسندگان

  • T Uchida
  • K Makita
  • Y Tsunoda
  • H Toyooka
  • K Amaha
چکیده

We evaluated the clinical performance of a continuous intraarterial blood gas monitoring (CIABG) system which includes a fluorometric intravascular sensor. Seventeen patients undergoing elective surgery were monitored perioperatively with the CIABG system (PB3300; Puritan Bennett, Carlsbad, CA). Conventional laboratory blood gas analyses (BGA) were performed simultaneously whenever indicated, and the values were compared with those obtained from the CIABG system. Monitoring time ranged from 24 to 72 hr. The biases (average error between PB3300 and BGA) of pH, PCO2 and PO2 were 0.003 pH unit, -2.8 mmHg, and 0.9 mmHg in the operating room (OR), and 0.005 pH unit, 3.9 mmHg, and 8.5 mmHg in the intensive care unit (ICU), respectively. The precision (standard deviation of the bias) of pH, PCO2 and PO2 were 0.030 pH unit, 2.1 mmHg, and 29.9 mmHg in the OR and 0.035 pH unit, 3.8 mmHg, and 14.7 mmHg in the ICU, respectively. Although the PB3300 system was clinically useful as a trend monitor, the system's precision and reliability were unacceptable for estimation of true blood gas values.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Design and Evaluation of a Pressure and Temperature Monitoring System for Pressure Ulcer Prevention

Introduction Pressure ulcers are tissue damages resulting from blood flow restriction, which occurs when the tissue is exposed to high pressure for a long period of time. These painful sores are common in patients and elderly, who spend extended periods of time in bed or wheelchair. In this study, a continuous pressure and temperature monitoring system was developed for pressure ulcer preventio...

متن کامل

A System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis

Background: Cardiac output (CO) is the total volume of blood pumped by the heart per minute and is a function of heart rate and stroke volume. CO is one of the most important parameters for monitoring cardiac function, estimating global oxygen delivery and understanding the causes of high blood pressure. Hence, measuring CO has always been a matter of interest to researchers and clinicians. Sev...

متن کامل

Evaluation of the Remler M2000 blood pressure recorder. Comparison with intraarterial blood pressure recordings both at hospital and at home.

The Remler M2000 is a semiautomated device that has been used to collect epidemiological data and assess blood pressure variability. It has been subjected to limited evaluation in operation, however, and no studies of its accuracy away from the hospital or office environment have been undertaken. We recruited a group of 28 patients with essential hypertension who were undergoing intraarterial a...

متن کامل

Gelsomino 1 Assessment of a Continuous Blood Gas Monitoring System in Animals during Circulatory stress. Short Title: Continuous Blood Gas Monitoring

Introduction: The study was aimed to determine the measurement accuracy of The CDI TM blood parameter monitoring system 500 (Terumo Cardiovascular Systems Corporation, Ann Arbor MI) in the real-time continuous measurement of arterial blood gases under different cardiocirculatory stress conditions Methods: Inotropic stimulation (Dobutamine 2.5 and 5 μg/kg/min), vasoconstriction (Argininevasopres...

متن کامل

Introducing intravascular microdialysis for continuous lactate monitoring in patients undergoing cardiac surgery: a prospective observational study

INTRODUCTION Lactate is a marker of hypoperfusion and may be used for risk assessment in critically ill patients. Although evidence suggests that repeated lactate measurements are of clinical interest, how and when lactate should be analyzed is controversial. Intravascular microdialysis provides a novel method for the continuous monitoring of lactate, which may be clinically beneficial in criti...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Canadian journal of anaesthesia = Journal canadien d'anesthesie

دوره 41 1  شماره 

صفحات  -

تاریخ انتشار 1994