Linear ion trap fabricated using rapid manufacturing technology.
نویسندگان
چکیده
This article presents the design, construction, and test results of a linear ion trap, fabricated using digital light processing (DLP), which is a low-cost 3D layer-based manufacturing technique. The ion trap was incorporated into a portable mass spectrometer system and experimental mass spectrum was obtained for methamphetamine (m/z 182), cocaine (m/z 304), and rhodamine B (m/z 443), with a maximum observed resolution (FWHM) of 260. For rhodamine B, tandem MS capability is also demonstrated. The mass range (and resolution at higher m/z) of the instrument is also demonstrated by spectrum obtained from Ultramark (m/z 1621). The spectra obtained for the DLP trap occur at a considerably lower rf voltage than a rectilinear ion trap of similar size, which is a consequence of the hyperbolic electrode geometry and, hence, smaller r(0) in the DLP case. High mass range with low voltage operation is especially important with regard to 'in the field' applications requiring low power consumption for extended periods of operation.
منابع مشابه
A Rapid, Simple, Liquid Chromatographic-Electrospray Ionization, Ion Trap Mass Spectrometry Method for the Determination of Finasteride in Human Plasma and its Application to Pharmacokinetic Study
A fast, accurate, sensitive, selective and reliable method using reversed-phase high performance liquid chromatography coupled to electrospray ionization ion trap mass spectrometry was developed and validated for the determination of finasteride in human plasma. After protein precipitation with perchloric acid, satisfactory separation was achieved on a Zorbax Eclipse® C8 analytical column u...
متن کاملA Rapid, Simple, Liquid Chromatographic-Electrospray Ionization, Ion Trap Mass Spectrometry Method for the Determination of Finasteride in Human Plasma and its Application to Pharmacokinetic Study
A fast, accurate, sensitive, selective and reliable method using reversed-phase high performance liquid chromatography coupled to electrospray ionization ion trap mass spectrometry was developed and validated for the determination of finasteride in human plasma. After protein precipitation with perchloric acid, satisfactory separation was achieved on a Zorbax Eclipse® C8 analytical column u...
متن کاملTransport of ions in a segmented linear Paul trap in printed-circuit-board technology
We describe the construction and operation of a segmented linear Paul trap, fabricated in printed-circuit-board technology with an electrode segment width of 500μm. We prove the applicability of this technology to reliable ion trapping and report the observation of Doppler-cooled ion crystals of 40Ca with this kind of trap. Measured trap frequencies agree with numerical simulations at the level...
متن کاملRapid Manufacturing of Co-Cr-Mo Implants by Three-Dimentional Printing Process for Orthopedic Applications
The fabrication of complex-shaped parts out of (wt %) Co-28Cr-6Mo alloy by three-dimensional printing (3DP) was studied using two grades of the alloy with average particle sizes of 20 and 75 μm. To produce sound specimens, 3DP processing parameters were tuned. The sintering behavior of the powders was characterized by the dilatometric analysis. Batch sintering in argon atmosphere at 1280 °...
متن کاملMiniaturized Linear Wire Ion Trap Mass Analyzer.
We report a linear ion trap (LIT) in which the electric field is formed by fine wires held under tension and accurately positioned using holes drilled in two end plates made of plastic. The coordinates of the hole positions were optimized in simulation. The stability diagram and mass spectra using boundary ejection were compared between simulation and experiment and good agreement was found. Th...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of the American Society for Mass Spectrometry
دوره 21 2 شماره
صفحات -
تاریخ انتشار 2010