Nuts & Bolts of Advanced Imaging: Theory and Demonstration for Coils, RF Shimming and SAR

نویسنده

  • Tamer S. Ibrahim
چکیده

Highlights:  At high frequency MRI operation and electrically large (≥ 7T head, ≥ 3T body), asymmetrical, inhomogeneous, and/or irregularly-shaped loading, integer multiples of phase-shifts and uniform amplitudes are not necessarily the ideal characteristics to impose on the voltages driving the transmit array/coil in order to obtain a homogeneous transmit (with safe electric field) and/or receive fields.  If none of the coils on the Tx Array is producing a good excitation in the ROI, PTX and/RF shimming will not work.  Image voids due lack of RF excitation are caused by inhomogeneities and/or unsuitable polarization of the electromagnetic fields.  At 7T, there is typically good intensity of RF magnetic field across the volume of human head. At many instances, it possess unsuitable polarization.  While the physics (Maxwell’s) is well established, it is really difficult to make up definitive rules regarding SARs (especially local SARs)

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Transmit and receive FDTD modeling as a valuable tool for RF coil development: validation of simulations with in vivo torso imaging at 7 Tesla

The ability to predict both transmit and receive characteristics in vivo is an important and necessary tool for the development of coils especially at ultra high magnetic fields (UHF). Accurate simulation permits the virtual development of coils saving both human and material resources [1, 2]. This is especially important during this period where large arrays and complex geometries are being pr...

متن کامل

Comparison of simulated parallel transmit body arrays at 3 T using excitation uniformity, global SAR, local SAR, and power efficiency metrics.

PURPOSE We compare the performance of eight parallel transmit (pTx) body arrays with up to 32 channels and a standard birdcage design. Excitation uniformity, local specific absorption rate (SAR), global SAR, and power metrics are analyzed in the torso at 3 T for radiofrequency (RF)-shimming and 2-spoke excitations. METHODS We used a fast cosimulation strategy for field calculation in the pres...

متن کامل

Assessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences

Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of  non-homogeneity in RF receive and transmit. This can be the most effective source of error in  quantitative  studies  in  MRI  imaging.  Part  of  this  non-homogeneity  demonstrates  the  characteristics of RF coil and part of it is due to the interaction of RF field with the material being  imaged...

متن کامل

Average SAR Constrained Local RF Shimming

Introduction Achieving given homogeneity and SAR constraints in RF shimming process is an important problem in high field MRI. There are different studies in the literature which dealt with the optimization of these parameters by using non-convex optimization techniques. In a previous work whole body SAR constrained RF shimming problem was solved in order to achieve whole slice transmit homogen...

متن کامل

Eigenmode Analysis of Transmit Coil Array for SAR-reduced B1 Mapping and RF shimming

Introduction The B1 transmit field inhomogeneity represents a serious problem in whole-body high field MRI (≥3T). B1 shimming based on measured B1 maps is a promising approach to cope with this problem and represents the primary application for parallel transmission at this point in time [1,2]. However, B1 mapping is still an error-prone and time consuming process [3], potentially resulting in ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2014