SU-E-T-168: Development of a Liquid Scintillation Detector for External Beam Dosimetry.
نویسندگان
چکیده
PURPOSE The goal of this research was to design a liquid scintillation dosimeter that could be used forrelative dosimetry of linear accelerator fields. The project emphasized minimization of cost and ease of use. METHODS The scintillator that was used in this research was BETAMAX- ES scintillation cocktail from MPBiomedical. This particular scintillator was selected due to its relatively high scintillation yield and lowcost. The entirety of the scintillator used the measurements was supplied free of cost. The housing for the liquid was constructed from PVC and is cylindrical with one tapered end. One fiber of the dual optical fibers transmits the generated photons to the CCD while the other fiber is used for Cerenkovsubtraction.The detector used comes from a Philips SPC880NC webcam. The plastic casing of the webcamwas removed so that only the printed circuit board, USB cable and lens eyepiece holder remained. Thesensor employed is the Sony ICX098QB CCD, which is 3.2mm by 2.4mm and each pixel is 5.6mm by 5.6mm. A small cylindrical insert was manufactured that was inserted into the lens eyepiece holder to get adequate mechanical coupling of the fibers to the CCD face. Images were acquired with a freeware image acquisition tool, SharpCap, and analyzed with theMatlab commercial math package from Mathworks. RESULTS Measurements have been performed that show that the detector is able to accurately measuretissue maximum ratio and the relative dose factor. The detector was able to accurately measurephysical wedge factors and made good predictions of the modulation factor for a patient's 7-field IMRT plan. CONCLUSIONS This work has shown that relative dosimetry can be performed using an inexpensive liquidscintillation detector. This could be expanded to include an array of liquid scintillator cells formeasurement of beam profiles and other more complex problems.
منابع مشابه
Performance characterization of a 3D liquid scintillation detector for discrete spot scanning proton beam systems.
Existing systems for proton beam dosimetry are limited in their ability to provide a complete, accurate, and detailed account of volumetric dose distribution. In this work, we describe the design and development of a portable, fast, and reusable liquid scintillator-based three-dimensional (3D) optical detection system for use in proton therapy. Our long-term goal is to use this system clinicall...
متن کاملWater-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements.
The properties of a new scintillation detector system for use in dosimetry of high-energy beams in radiotherapy have been measured. The most important properties of these detectors are their hgh spatial resolution and their nearly water-equivalence. Measurements have shown that they have excellent reproducibility and stability, and a linear response versus dose-rate. It is also shown that they ...
متن کاملWater-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration.
A minimally perturbing plastic scintillation detector has been developed for the dosimetry of high-energy beams in radiotherapy. The detector system consists of two identical parallel sets of radiation-resistant optical fibre bundles, each connected to independent photomultiplier tubes (PMTs). One fibre bundle is connected to a miniature water equivalent plastic scintillator and so scintillatio...
متن کاملDetermination of chemical and radiochemical purity of [14C] clozapine by high performance liquid chromatography and liquid scintillation [Persian]
Introduction: The mechanism of action and metabolism of clozapine, a complex drug used in psychopharmacology, is not completely understood. The effects of this compound in human body are related to some interactions with serotonin and dopamine neurotransmitter receptors. To further elucidate the mechanism of action and to support ongoing drug metabolism studies, [14C] clozapine with suita...
متن کاملThe three-dimensional scintillation dosimetry method: test for a 106Ru eye plaque applicator.
The need for fast, accurate and high resolution dosimetric quality assurance in radiation therapy has been outpacing the development of new and improved 2D and 3D dosimetry techniques. This paper summarizes the efforts to create a novel and potentially very fast, 3D dosimetry method based on the observation of scintillation light from an irradiated liquid scintillator volume serving simultaneou...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Medical physics
دوره 39 6Part12 شماره
صفحات -
تاریخ انتشار 2012