Impaction bone grafting for tibial defects in knee replacement surgery. Results at two years.
نویسندگان
چکیده
Bone loss with large defects poses a complex and challenging problem in primary and revision knee arthroplasty. The defects are often irregular and difficult to quantify. One of the techniques available to restore bone in such cases is Knee Impaction Bone Grafting (KIBG); however, the clinical literature to support this technique is weak. Since 2006 we have used impaction bone grafting for contained and uncontained large tibial defects in primary and revision total knee arthroplasty. We have prospectively studied 11 patients with large tibial defects treated at the Exeter Knee Reconstruction Unit with KIBG using a short cemented stem following the Slooff-Ling philosophy. Average age was 66 years (41-86 years). Minimum follow-up was 2 years. The Knee Society Scores improved from 27.4 to 89.2 on average, with Knee Society Function score and WOMAC increasing by 263 and 23.2 points respectively. The mean post-operative flexion was 112 degrees. The average gain in motion over preoperative value was 20 degrees. At two years there were no mechanical failures. None of the patients have required secondary procedures or further revisions. All radiographs showed incorporation and remodelling of the graft. The only complication was a superficial dysaesthesia around the operative scar. Although being time consuming and technically demanding, KIBG for substantial tibial bone loss has shown excellent versatility and good short term results, providing a stable construct with immediate weight bearing post operatively. In view of previous concerns regarding early incorporation and stability of impaction bone grafting in the tibia, we present our early results at 2 years. This technique has become our preferred technique for treating substantial bone loss in tibial defects seen in primary and revision knee arthroplasty surgery.
منابع مشابه
Filling of segmental bone defects in revision knee arthroplasty using morsellized bone grafts contained within a metal mesh.
For revision knee surgery with uncontained tibial bone defects, the authors report the containment of compacted morsellized allograft using metal-wire mesh, followed by implantation of a cemented total knee prosthesis. This method is comparable to the "impaction grafting technique" described for revision hip surgery and could be an alternative to metal wedges, augmented components, custom-made ...
متن کاملA Comparative Biomechanical Analysis
J Bone Joint Surg [Br] 2009;91-B:271-7. Received 16 June 2008; Accepted after revision 26 September 2008 The treatment of bony defects of the tibia at the time of revision total knee replacement is controversial. The place of compacted morsellised bone graft is becoming established, particularly in contained defects. It has previously been shown that the initial stability of impaction-grafted t...
متن کامل10-year results following impaction bone grafting of major bone defects in 29 rotational and hinged knee revision arthroplasties
BACKGROUND AND PURPOSE Substantial bone loss in revision total knee arthroplasty (TKA) is a challenging problem. We studied whether impaction bone grafting provides long-term restoration of bone stock in the treatment of major bone defects in revision surgery of rotational and hinged knee arthroplasties (LINK Endo-Model). PATIENTS AND METHODS Between 1996 and 2006, 29 knees in 29 patients und...
متن کاملComputer-assisted revision total knee replacement.
A technique for performing allograft-augmented revision total knee replacement (TKR) using computer assistance is described, on the basis of the results in 14 patients. Bone deficits were made up with impaction grafting. Femoral grafting was made possible by the construction of a retaining wall or dam which allowed pressurisation and retention of the graft. Tibial grafting used a mixture of cor...
متن کاملMechanical testing of impaction bone grafting in the tibia: initial stability and design of the stem.
Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray. We developed a system for ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Acta orthopaedica Belgica
دوره 79 2 شماره
صفحات -
تاریخ انتشار 2013