Reduction of the radiofrequency heating of metallic devices using a dual-drive birdcage coil.

نویسندگان

  • Yigitcan Eryaman
  • Esra Abaci Turk
  • Cagdas Oto
  • Oktay Algin
  • Ergin Atalar
چکیده

In this work, it is demonstrated that a dual-drive birdcage coil can be used to reduce the radiofrequency heating of metallic devices during magnetic resonance imaging. By controlling the excitation currents of the two channels of a birdcage coil, the radiofrequency current that is induced near the lead tip could be set to zero. To monitor the current, the image artifacts near the lead tips were measured. The electric field distribution was controlled using a dual-drive birdcage coil. With this method, the lead currents and the lead tip temperatures were reduced substantially [<0.3 °C for an applied 4.4 W/kg SAR compared to >4.9 °C using quadrature excitation], as demonstrated by phantom and animal experiments. The homogeneity of the flip angle distribution was preserved, as shown by volunteer experiments. The normalized root-mean-square error of the flip angle distribution was less than 10% for all excitations. The average specific absorption rate increased as a trade-off for using different excitation patterns.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Safety of metallic implants in magnetic resonance imaging.

Magnetic resonance (MR) imaging has become a commonly accepted medical procedure. Manufacturers of medical implant devices are submitting claims that their devices are safe and effective in a MR environment. This paper concentrates on the issue of heating of patients due to the interaction of metallic implants with the strong radiofrequency (RF) magnetic field produced by the MR scanner. The co...

متن کامل

Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging

Deep Brain Stimulation (DBS) is increasingly used to treat a variety of brain diseases by sending electrical impulses to deep brain nuclei through long, electrically conductive leads. Magnetic resonance imaging (MRI) of patients pre- and post-implantation is desirable to target and position the implant, to evaluate possible side-effects and to examine DBS patients who have other health conditio...

متن کامل

RADIOFREQUENCY COIL DESIGN AND APPLICATION TO MAGNETIC RESONANCE IMAGING AND CONTROL OF MICRO-BEADS by

Radio Frequency coil development, one component of a functioning magnetic resonance imaging (MRI) system, has evolved since the beginning of MRI into a very diverse area. Today for example, transmit and receive coils, volume and surface coils, and single channel and multiple channel coils are all included. Coils are created to fit the particular needs of specific research purposes, taking into ...

متن کامل

Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil

The transmit-receive (Tx/Rx) birdcage head coil is often used for excitation instead of the body coil because of the presumably lower risk of heating in and around conductive implants. However, this common practice has not been systematically tested. To investigate whether the Tx/Rx birdcage head coil produces less heating than the body coil when scanning individuals with implants, we used a 3T...

متن کامل

Reduction of implant RF heating through modification of transmit coil electric field.

In this work, we demonstrate the possibility to modify the electric-field distribution of a radio frequency (RF) coil to generate electric field-free zones in the body without significantly altering the transmit sensitivity. Because implant heating is directly related to the electric-field distribution, implant-friendly RF transmit coils can be obtained by this approach. We propose a linear bir...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Magnetic resonance in medicine

دوره 69 3  شماره 

صفحات  -

تاریخ انتشار 2013