Evaluation of the PIMA point-of-care CD4 analyzer in VCT clinics in Zimbabwe.
نویسندگان
چکیده
Point-of-care (POC) CD4 testing was implemented at a stand-alone HIV voluntary testing and counseling centre in Harare, Zimbabwe. To validate the use of this new technology, paired blood samples were collected from 165 patients either by a nurse or a laboratory technician and tested using POC and conventional laboratory CD4 machines. Finger prick (capillary) blood was collected directly into the PIMA POC CD4 Analyzer cartridges and tested immediately, whereas venous blood collected into evacuated tubes was used for CD4 enumeration on a Becton Dickinson FACSCalibur. There was no significant difference in mean absolute CD4 counts between the POC PIMA and Becton Dickinson FACSCalibur platforms (+7.6 cells/microL; P = 0.72). Additionally, there was no significant difference in CD4 counts between the platforms when run by either a nurse (+18.0 cells/microL; P = 0.49), or a laboratory technicians (-3.1 cells/microL; P = 0.93). This study demonstrates that POC CD4 testing can be conducted in a voluntary testing and counseling setting for staging HIV-positive clients. Both nurses and laboratory technicians performed the test accurately, thereby increasing the human resources available for POC CD4 testing. By producing same-day results, POC CD4 facilitates immediate decision-making, patient management and referral and may help improve patient care and retention. POC CD4 may also alleviate testing burdens at traditional central CD4 laboratories, hence improving test access in both rural and urban environments.
منابع مشابه
PIMA Point of Care CD4+ Cell Count Machines in Remote MNCH Settings: Lessons Learned from Seven Districts in Zimbabwe
An evaluation was commissioned to generate evidence on the impact of PIMA point-of-care CD4+ count machines in maternal and new-born child health settings in Zimbabwe; document best practices, lessons learned, challenges, and recommendations related to scale up of this new technology. A mixed methodology approach that included 31 in-depth interviews with stakeholders involved in procurement, di...
متن کاملField Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
OBJECTIVE To evaluate the accuracy of the PIMA point-of-care CD4 analyzer (PIMA) under field conditions in comparison to the current CD4 count system (FACSCalibur), and to evaluate the operational suitability and acceptability of health professionals (HP) and HIV-patients in using the PIMA in health clinics in the Amazon Region. METHODS CD4 counts were measured onsite by the PIMA using finger...
متن کاملUtility of the point of care CD4 analyzer, PIMA, to enumerate CD4 counts in the field settings in India
UNLABELLED BACKGROUND In resource limited settings non-availability of CD4 count facility at the site could adversely affect the ART roll out programme. Point of care CD4 enumerating equipments can make the CD4 count available at the site of care and improve the patients' management considerably. This study is aimed at determining the utility of a Point of Care PIMA CD4 analyzer (Alere, Germ...
متن کاملEvaluation of PIMA Point-of-care CD4 Analyzer in Yunnan, China
BACKGROUND CD4 count is used to determine antiretroviral therapy (ART) eligibility. In China, flow cytometers are mostly located in urban areas with limited access by patients residing in remote areas. In an attempt to address this issue, we conducted a study to validate the performance of Alere PIMA point-of-care CD4 analyzer. METHODS Venous and finger-prick blood specimens were collected fr...
متن کاملField Evaluation of PIMA Point-of-Care CD4 Testing in Rakai, Uganda
OBJECTIVE To assess the accuracy of PIMA Point-of-Care (POC) CD4 testing in rural Rakai, Uganda. METHODS 903 HIV positive persons attending field clinics provided a venous blood sample assessed on site using PIMA analyzers per manufacturer's specifications. The venous samples were then run on FACSCalibur flow cytometry at a central facility. The Bland-Altman method was used to estimate mean b...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of acquired immune deficiency syndromes
دوره 55 1 شماره