Quantification and minimization of magnetic susceptibility artifacts on GRE images.

نویسندگان

  • J D Port
  • M G Pomper
چکیده

PURPOSE The purpose of this work was to determine the optimal imaging parameters for minimization of metallic susceptibility artifacts during gradient echo (GRE) imaging. METHOD We performed GRE imaging of titanium screws in a nickel-doped agarose gel phantom, systematically varying several parameters to characterize and quantify susceptibility artifacts. RESULTS The greatest reduction in artifact size came from using a short TE; increasing the frequency matrix and decreasing the slice thickness also contributed substantially to reducing the artifact size. Whenever possible, implanted prostheses should be aligned with the main magnetic field to minimize artifact size. Parameters with negligible effect on artifact size included bandwidth, phase encode matrix, and field of view. CONCLUSION Radiologists can easily adjust the above parameters in their imaging protocols to improve GRE image quality in patients with implanted metallic devices.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Non-invasive quantification of liver fat content by different Gradient Echo MRI sequences in patients with Non-Alcoholic Fatty Liver Disease (NAFLD)

Introduction: Non-invasive quantification of liver fat by Gradient echo (GRE) Technique is an interesting issue in quantitative MRI. Despite the numerous advantages of this technique, fat measurement maybe biased by confounding and effects. The aim of this study was to evaluate the GRE pulse sequences with different   and  weighting for liver fat quantification in patients with...

متن کامل

[Study of aliasing ERROR with SENSE in body diffusion image using single-shot EPI at 3T].

Magnetic field inhomogeneity causes artifacts in MRI. For example, in single-shot echo-planar imaging (EPI), they often appear as severe geometric distortions along the phase-encoding direction. Sensitivity encoding (SENSE) is useful in reducing the distortion in EPI since it only acquires partial k-space data using multiple receiver channels. In SENSE, a reference scan usually needs to be perf...

متن کامل

3D SAP-EPI motion-corrected fast susceptibility weighted imaging

Introduction: Susceptibility-weighted imaging (SWI) has been utilized as a useful contrast mechanism in MRI that accentuates the paramagnetic properties of blood products (1,2). With the use of both magnitude and phase images, SWI can provide improved conspicuity of venous blood vessels and other sources of susceptibility effects (2). Typically, the SWI acquisition uses a high-resolution, three...

متن کامل

Improved susceptibility weighted imaging method using multi-echo acquisition.

PURPOSE To introduce novel acquisition and postprocessing approaches for susceptibility weighted imaging (SWI) to remove background field inhomogeneity artifacts in both magnitude and phase data. METHODS The proposed method acquires three echoes in a three-dimensional gradient echo (GRE) sequence, with a field compensation gradient (z-shim gradient) applied to the third echo. The artifacts in...

متن کامل

WEEKDAY EDUCATIONAL COURSE Everything You Wanted to Know about Magnetic Susceptibility & Why It Is Important

Gradient Recalled Echo (GRE) based MR imaging techniques have wide applications in MRI. Due to the absence of refocusing RF pulses GRE sequences are fast and since low flip angles are usually applied, GRE sequences have low specific absorption rate, hence are particularly well suited for high field MRI. While MR signal generated from GRE techniques is complex, most traditional applications rely...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of computer assisted tomography

دوره 24 6  شماره 

صفحات  -

تاریخ انتشار 2000