The valve-in-valve technique for treatment of aortic bioprosthesis malposition an analysis of incidence and 1-year clinical outcomes from the italian CoreValve registry.
نویسندگان
چکیده
OBJECTIVES We appraised the incidence and clinical outcomes of patients who were treated with the valve-in-valve (ViV) technique for hemodynamically destabilizing paraprosthetic leak (PPL). BACKGROUND Device malpositioning causing severe PPL after transcatheter aortic valve implantation is not an uncommon finding. It occurs after release of the prosthesis, leading to hemodynamic compromise. It can be managed successfully in selected cases with implantation of a second device inside the malpositioned primary prosthesis (ViV technique). METHODS Consecutive patients (n = 663) who underwent transcatheter aortic valve implantation with the 18-F CoreValve ReValving System (Medtronic, Inc., Minneapolis, Minnesota) at 14 centers across Italy were included in this prospective web-based registry. We identified patients treated with the ViV technique for severe PPL and analyzed their clinical and echocardiographic outcomes. Primary end points were major adverse cerebrovascular and cardiac events and prosthesis performance at the 30-day and midterm follow-up. RESULTS Overall procedural success was obtained in 650 patients (98.0%). The ViV technique was used in 24 (3.6%) of 663 patients. The 30-day major adverse cerebrovascular and cardiac event rates were 7.0% and 0% in patients undergoing the standard procedure and ViV technique, respectively (p = 0.185); the mortality rates were 5.6% versus 0% in patients undergoing the standard procedure and ViV technique, respectively (p = 0.238). There was an improvement in the mean transaortic gradient in all patients without significant difference between the 2 groups (from 52.1 ± 17.1 mm Hg and 45.4 ± 14.8 mm Hg [p = 0.060] to 10.1 ± 4.2 mm Hg and 10.5 ± 5.2 mm Hg, respectively [p = 0.838]). At 12 months, the major adverse cerebrovascular and cardiac event rates in the standard procedure and ViV technique groups were 4.5% and 14.1%, respectively (p = 0.158), and the mortality rates were 4.5% versus 13.7%, respectively (p = 0.230). CONCLUSIONS This large, multicenter registry provides important information about the feasibility, safety, and efficacy of the ViV technique with the third-generation CoreValve ReValving System. The clinical and echocardiographic end points compare favorably with those of patients undergoing the standard procedure. The ViV technique offers a viable therapeutic option in patients with acute significant PPL without recourse to emergent surgery.
منابع مشابه
Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results from the global valve-in-valve registry.
BACKGROUND Transcatheter aortic valve-in-valve implantation is an emerging therapeutic alternative for patients with a failed surgical bioprosthesis and may obviate the need for reoperation. We evaluated the clinical results of this technique using a large, worldwide registry. METHODS AND RESULTS The Global Valve-in-Valve Registry included 202 patients with degenerated bioprosthetic valves (a...
متن کاملMatched Comparison of Self-Expanding Transcatheter Heart Valves for the Treatment of Failed Aortic Surgical Bioprosthesis: Insights From the Valve-in-Valve International Data Registry (VIVID).
BACKGROUND Transcatheter valve-in-valve implantation is an established therapy for high-risk patients with failed surgical aortic bioprosthesis. There are limited data comparing outcomes of valve-in-valve implantation using different transcatheter heart valves (THV). METHODS AND RESULTS Patients included in the Valve-in-Valve International Data registry (VIVID) and treated with self-expanding...
متن کاملMatched Comparison of Self-Expanding Transcatheter Heart Valves for the Treatment of Failed Aortic Surgical Bioprosthesis
Background—Transcatheter valve-in-valve implantation is an established therapy for high-risk patients with failed surgical aortic bioprosthesis. There are limited data comparing outcomes of valve-in-valve implantation using different transcatheter heart valves (THV). Methods and Results—Patients included in the Valve-in-Valve International Data registry (VIVID) and treated with selfexpanding TH...
متن کاملProsthesis-patient mismatch after transcatheter aortic valve implantation with the Medtronic-Corevalve bioprosthesis.
AIMS Prosthesis-patient mismatch (P-PM) is an important determinant of morbidity and mortality following open aortic valve replacement. The aims of this study were to report its incidence and determinants following transcatheter aortic valve implantation (TAVI) with the Corevalve bioprosthesis, which have-thus far-not been described. METHODS AND RESULTS Patients with severe calcific aortic st...
متن کاملTranscatheter Aortic Valve Replacement: A Review Article
Transcatheter aortic valve replacement (TAVR) is a novel therapeutic intervention for the replacement of severely stenotic aortic valves in high-risk patients for standard surgical procedures. Since the initial PARTNER trial results, use of TAVR has been on the rise each year. New delivery methods and different valves have been developed and modified in order to promote the minimally invasive p...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of the American College of Cardiology
دوره 57 9 شماره
صفحات -
تاریخ انتشار 2011