Tackling matrix effects during development of a liquid chromatographic-electrospray ionisation tandem mass spectrometric analysis of nine basic pharmaceuticals in aqueous environmental samples.
نویسندگان
چکیده
When developing an LC-MS/MS-method matrix effects are a major issue. The effect of co-eluting compounds arising from the matrix can result in signal enhancement or suppression. During method development much attention should be paid to diminish matrix effects as much as possible. The present work evaluates matrix effects from aqueous environmental samples in the simultaneous analysis of a group of nine specific pharmaceuticals with LC-ESI/MS/MS: flubendazole, propiconazole, pipamperone, cinnarizine, ketoconazole, miconazole, rabeprazole, itraconazole and domperidone. Solutions to diminish signal suppression were examined: optimisation of the sample preparation, decrease of the flow rate, and the use of appropriate internal standards. Several SPE-stationary phases were tested in view of retention of the analytes: Oasis HLB, C8, Phenyl, Strata X-polymer RP sorbent and Strata X-polymeric SCX/RP sorbent. Oasis HLB showed the best retention for all analytes. The Oasis HLB SPE-method was optimised, but analyses showed high matrix suppression. Therefore, a second SPE-method, on a phenyl stationary phase (the second best option), was also optimised. A comparison of the matrix effect was made between the two procedures: the phenyl-method was less subject to matrix effects, however, the average matrix effect (ME%) of 46% indicated that matrix effects where still present. Several optimisation options for the phenyl-method were evaluated: addition of a ferric nitrate solution before extraction, application of an alkaline wash step, and use of a second SPE-cartridge, either a NH2-column or a florisil column. A more efficient sample clean-up was achieved by applying the extract after extraction on the phenyl column and after dilution with chloroform, onto a NH2-column (average ME%: 53%). In addition, applying a post-column split (1:5), further reduced matrix effects (average ME%: 65%). Labelled internal standards are the best way to tackle matrix effects, but no such internal standards were commercially available for the analytes of interest. The thorough search and application of four internal standards (structural analogues) was beneficial and compensates the matrix effect partially (average ME%: 83%). In an attempt to reduce the analysis time Speedisk phenyl columns were applied. Under these conditions matrix effects decreased even more while recoveries were between 91 and 109%. Different kinds of surface water samples were analyzed, and different matrix effects were observed. For this reason, standard addition will be used to perform quantitative analysis.
منابع مشابه
Liquid Chromatography Tandem Mass/Mass Spectrometry for the Quantification of Fudosteine in Human Serum without Precolumn Derivatization
A quantitative analysis method for fudosteine in human serum by high-performance liquid chromatography-electrospray ionization tandem mass spectrometry (HPLC-ESI/MS/MS) was established, which shows high sensitivity and selectivity. The mobile phase composition was 75% 20 mM acetic acid and 25% acetonitril, which was pumped at a flow rate of 0.40 mL/min. The overall chromatographic run time was ...
متن کاملLiquid Chromatography Tandem Mass/Mass Spectrometry for the Quantification of Fudosteine in Human Serum without Precolumn Derivatization
A quantitative analysis method for fudosteine in human serum by high-performance liquid chromatography-electrospray ionization tandem mass spectrometry (HPLC-ESI/MS/MS) was established, which shows high sensitivity and selectivity. The mobile phase composition was 75% 20 mM acetic acid and 25% acetonitril, which was pumped at a flow rate of 0.40 mL/min. The overall chromatographic run time was ...
متن کاملUsing calibration approaches to compensate for remaining matrix effects in quantitative liquid chromatography/electrospray ionization multistage mass spectrometric analysis of phytoestrogens in aqueous environmental samples.
Signal suppression is a common problem in quantitative liquid chromatography/electrospray ionization multistage mass spectrometric (LC/ESI-MS(n)) analysis in environment samples, especially in highly loaded wastewater samples with highly complex matrix. Optimization of sample preparation and improvement of chromatographic separation are prerequisite to improve reproducibility and selectivity. M...
متن کاملAnalysis of hexabromocyclododecane diastereomers and enantiomers by liquid chromatography/tandem mass spectrometry: chromatographic selectivity and ionization matrix effects.
Hexabromocyclododecane (HBCD) is a flame retardant that is undergoing environmental risk assessment. The liquid chromatographic retention and electrospray ionization matrix effects were investigated for HBCD methods of analysis for environmental matrices. Column selectivity towards HBCD diastereomers was evaluated for C30 and C18 stationary phases under different mobile phase conditions and col...
متن کاملComparative Pharmacokinetics of Berberine, Palmatine and Jatrorrhizine in Rat Plasma after Oral Administration of Rhizoma coptidis and Zuojinwan Using Liquid Chromatography-Tandem Mass Spectrometry
A selective and sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed for the determination of berberine, palmatine and jatrorrhizine in rat plasma. Target compounds, together with the internal standard (metronidazole), were extracted from rat plasma samples by protein precipitation with acetonitrile-methanol (1:2, v/v). Chromatography was carried out using a ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of chromatography. A
دوره 1123 1 شماره
صفحات -
تاریخ انتشار 2006