Validation of an automated technique for determining the mechanical characteristics of coronary arteries during balloon angioplasty: laboratory assessment with necropsy segments.
نویسندگان
چکیده
OBJECTIVES To develop a technique for automatic inflation of a percutaneous transluminal coronary angioplasty (PTCA) balloon, with continuous measurement of the balloon pressure and volume; to validate the technique for determining the mechanical characteristics of coronary arteries. METHODS During necropsy examination of the hearts of nine patients, 17 coronary artery samples were obtained for histological examination. A PTCA balloon was inserted into each artery, and the balloon pressure and volume were measured continuously during four repeat automatic inflations of the balloon. RESULTS Of the 17 arteries, eight showed elastic, six plastic, and three fracture pressure-volume deformation characteristics. For the plastic deformations, the first inflation required a higher pressure than subsequent repeat inflations of 82 (61) kPa (mean (SD), range 25 to 175 kPa). For the three in the fracture group, the pressure drop because of the fracture occurred between 210 and 540 kPa. Two of these three showed a tear on visual inspection, and the other showed disruption of the intimal plaque on blinded histological examination. Of the six with plastic deformation characteristics alone, one showed a tear, and on histological examination two others showed splitting of the internal and external elastic lamina and one showed separation of intima and media. None in the elastic group showed any of these characteristics. CONCLUSIONS Plastic and fracture deformation characteristics could be differentiated from elastic characteristics. Visual or histological evidence of fracturing was present in all three arteries identified during angioplasty as having pressure-volume fracture characteristics.
منابع مشابه
Immediate and Short-term Follow-Up of Aortic Coarctation Balloon Angioplasty and Stenting
Background Aortic Coarcatation (CoA) is one of the congenital heart diseases with the rate of 5-8% of Coronary heart diseases(CHDs). Balloon angioplasty is now one of the effective way of treatment for CoA, native or Re-coarctation (Re-CoA). We aimed to assess the immediate, and short term response to angioplasty and stenting, and also complications. Materials and Methods Balloon angioplasty wi...
متن کاملClinical Implications of Congenital Absence of Circumflex Coronary Artery
Introduction: Coronary artery anomalies are rare clinical entities reported in 0.6% to 5.6% of diagnostic coronary angiographies. Anomalous origins of coronary arteries from distal segments are rarely reported. Presented herein is a coronary anomaly in which the circumflex artery (CX) originated as a continuation of the posterolateral branch of the right coronary artery (RCA) w...
متن کاملDigital subtraction angiography in angioplasty of total coronary artery occlusion.
This report describes a digital subtraction angiographic technique used to simultaneously display the proximal and distal segments of three totally occluded coronary arteries in two patients undergoing percutaneous transluminal coronary angioplasty (PTCA). The advantage of this technique over routine cineangiography for PTCA of total occlusions is illustrated.
متن کاملEvaluating long-term outcomes of coronary angioplasty with or without post-dilatation
Introduction: Post-dilatation is associated with a simultaneous expansion of the stents that enhances the angioplasty outcomes. However, increased risk of mortality and morbidity has been reported in patients with acute myocardial infarction (AMI) which has provoked considerable controversies concerning its efficiency. Materials and Methods: During a two-...
متن کاملAssessment of left ventricular performance during percutaneous transluminal coronary angioplasty: a study by intravenous digital subtraction ventriculography.
Left ventricular performance during percutaneous transluminal coronary angioplasty was assessed in 52 patients by intravenous digital subtraction ventriculography. After injection of contrast into the right atrium ventriculograms were obtained before and during balloon inflation. In 37 patients they were also obtained after the procedure. A 12 lead electrocardiogram was monitored throughout. Du...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Heart
دوره 79 6 شماره
صفحات -
تاریخ انتشار 1998