"Delfia" and "Amerlite": two sensitive nonisotopic immunoassay systems for assay of thyrotropin compared.

نویسندگان

  • A J Parnham
  • I F Tarbit
چکیده

We assessed the LKB "Delfia" (time-resolved dissociation-enhanced lanthanide fluoroimmunoassay) and the Amersham "Amerlite" (enhanced luminescent immunometry) assays of thyrotropin in serum. Both assays are sensitive (respective detection limits: 0.02 and 0.04 milli-int. unit/L) and have very good within- and between-batch precision over a wide range of thyrotropin concentrations. Results by the two methods correlate well (r = 0.992); the regression equation is: Amerlite = 0.915 Delfia - 0.33 milli-int. unit/L. The standard curve for the Delfia assay was linear, but that for the Amerlite assay showed some deviation from linearity below 0.5 milli-int. unit/L. Both assays have a negative bias in comparison with radiolabeled immunoradiometric assays, as judged by results for samples from the Quality Assurance Scheme. Both assays discriminate well between hyper-, hypo-, and euthyroid subjects, and results for thyrotropin for most patients with nonthyroidal illness were within the euthyroid reference interval. Both assays are convenient to perform and are based on systems that provide a viable alternative to radioimmunoassay.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Four nonisotopic immunoassays of free thyroxin evaluated.

We evaluated four new nonisotopic immunoassays of free thyroxin (FT4)--Amerlite FT4 (Amersham International), Magic Lite FT4 (Ciba Corning Diagnostics), Stratus FT4 (Dade--Baxter Travenol), and FT4 Enzelsa (Compagnie ORIS Industrie)--by comparison with two FT4 radioimmunoassays: Amerlex and Sclavo. Inter- and intra-assay coefficients of variation were less than 10% in the working range and ther...

متن کامل

Chemiluminescence immunoassay of thyrotropin with acridinium-ester-labeled antibody evaluated and compared with two other immunoassays.

A new chemiluminometric immunoassay of thyrotropin (TSH) involves antibody labeled with acridinium ester ("Magic Lite System," Ciba Corning Diagnostic Corp.). The assay is rapid, with two incubations totaling 2.5 h, requires two standards per run, and takes 10 s per sample for the quantification step. Analytical performance, within- and between-run reproducibilities, and linearity were excellen...

متن کامل

Four sensitive thyrotropin assays critically evaluated and compared.

We compared four sensitive procedures for thyrotropin (TSH)--Corning's Magic-Lite, ElectroNucleonics' Delfia, Baxter's Stratus, Nichols' Allegro--for their ability to completely discriminate TSH concentrations in sera in euthyroidism, hyperthyroidism, and hypothyroidism. We evaluated the analytical and clinical performance of these procedures according to previously published criteria. All proc...

متن کامل

Clinical and laboratory studies of time-resolved fluorescence immunoassays of thyrotropin and free triiodothyronine.

We evaluated a new time-resolved immunofluorometric assay (TR-IFMA) for thyrotropin (TSH) and a time-resolved fluoroimmunoassay (TR-FIA) for free triiodothyronine (FT3) in serum. Delfia hTSH Ultra is a solid-phase two-site TR-IFMA involving three monoclonal antibodies directed against different antigenic sites on the human TSH molecule. It offers improved analytical (+3 SD of zero) and function...

متن کامل

Enzyme immunoassays for IgG and IgM antibodies to Toxoplasma gondii based on enhanced chemiluminescence.

AIMS To evaluate the clinical performance of enzyme immunoassays for IgG and IgM antibodies to Toxoplasma gondii based on enhanced chemiluminescence. METHODS Classification of routine clinical samples from the originating laboratories was compared with that obtained using the chemiluminescence based assays. Resolution of discordant results was achieved by testing in alternative enzyme immunoa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 33 8  شماره 

صفحات  -

تاریخ انتشار 1987