Comparing and evaluating the efficacy of the TOR18FG Leeds test X-ray phantom for T-rays.
نویسندگان
چکیده
The commercially available X-ray fluoroscopy quality assurance phantom, the Leeds test object TOR18FG, was found to be suitable to assess T-ray image quality in the range (0.1-0.4) THz at a depth of 0.5 cm. Previous to this only custom made phantoms, made especially for the T-ray region, assessed T-ray spatial resolution. However, if sub-wavelength techniques are used, the Leeds test phantom may be implemented to measure the T-ray systems spatial resolution, allowing us to directly compare X-ray and T-ray spatial resolution. The systems compared include a Gulmay Orthovoltage machine (X-ray), the On Board Imager (OBI) of a Varian linear accelerator (X-ray), a two-colour system (T-ray) and Terahertz Time Domain Spectroscopy (THz-TDS) system. X-rays were found to have a spatial resolution of 1.25 lp/mm using the On Board Imager of a Varian Linear Accelerator whilst T-rays imaged using a broadband source imaged through a spatial pinhole had a spatial resolution of 0.56 lp/mm. The TOR18FG background material was found to block, 90% and 99% of the broadband T-rays emitted from a THz-TDS photo-conductive emitter, at 0.4 THz and 0.53 THz respectively. Contrast sensitivity was found to be 3% for 25 cm × 25 cm X-ray field at 65 kV, whilst this value could not be established for T-rays using the TOR18FG. All contrast circles were found to be the same for T-rays i.e. all 40% at 0.1 THz. Images of the same leaf were taken with diagnostic X-rays and both broadband and continuous wave (CW) T-ray systems. T-rays proved superior in providing image contrast, for a hydrated leaf, over X-rays.
منابع مشابه
A Monte Carlo Study on Dose Enhancement by Homogeneous and Inhomogeneous Distributions of Gold Nanoparticles in Radiotherapy with Low Energy X-rays
Background: To enhance the dose to tumor, the use of high atomic number elements has been proposed.Objective: The aim of this study is to investigate the effect of gold nanoparticle distribution on dose enhancement in tumor when the tumor is irradiated by typical monoenergetic X-ray beams by considering homogeneous and inhomogeneous distributions of gold nanoparticles (GNPs) in the tumor.Method...
متن کاملComparing X-ray dose reduction capability of silicon-bismuth micro- and nanocomposite shields using chest CT test
Comparison of radiation shields of different thicknesses and structures (especially different particle sizes of their constituents) is helpful to select an appropriate shield for patient protection in medical imaging tests particularly CT-scan. In this study micro- and nano-powders of bismuth were mixed with silicone matrix to make composite radiation shields for breast. The shields were provid...
متن کاملEvaluation of depth dose characteristics of superficial X-rays machine using different kVp and applicators diameter
Background: Radiotherapy Treatment Planning requires different dosimetric quantities as input in order to calculate a desired dose distribution. This study has been focused to evaluate the depth dose characteristics of superficial X-rays being used for radiotherapy treatment. Materials and Methods: Computerized 3-D water phantom of multi-data system was used. The measurements were made...
متن کاملComparison of the photon charge between water and solid phantom depending on depth
Background: This study assessed the clinical usefulness of the solid phantom, which may compensate for the disadvantages of the water phantom, by comparing the radiological doses between the two depending on their depths. Materials and Methods: The experimental equipment used was a linear accelerator for medical use, water phantom, solid phantom, Farmer type ion chamber and electrometer. The di...
متن کاملQualitative Evaluation of Digital Hand X-rays is Not a Reliable Method to Assess Bone Mineral Density
Object: The gold standard for evaluating bone mineral density is dual energy x-ray absorptiometry (DEXA). Prior studies have shown poor reliability using analog wrist X-rays in diagnosing osteoporosis. Our goal was to investigate if there was improved diagnostic value to visual assessment of digital hand X-rays in osteoporosis screening. We hypothesized that similar to analog counterparts, dig...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Quantitative imaging in medicine and surgery
دوره 3 1 شماره
صفحات -
تاریخ انتشار 2013