Dialysis catheter fibrin sheath stripping: a useful technique after failed catheter exchange

نویسندگان

  • AF Mohamad Ali
  • E Uhwut
  • SK Liew
چکیده

Fibrin sheath formation around long-term haemodialysis catheter is a common cause of failed dialysis access. Treatment options include pharmacological and mechanical methods. This paper reports a case of failed dialysis access due to fibrin sheath encasement. Pharmacologic thrombolysis, mechanical disruption using guide wire and catheter exchange had failed to address the issue. Eventually, fibrin sheath stripping using the loop snare technique was able to successfully restore the catheter function.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Percutaneous fibrin sleeve stripping of failing haemodialysis catheters.

BACKGROUND One of the most frequent reasons for failure of haemodialysis lines is catheter blockage caused by fibrin sheath formation. We report our experience of percutaneous fibrin sheath stripping in treating this problem. METHODS A consecutive series of failing haemodialysis catheters underwent percutaneous fibrin sheath stripping in an attempt to retrieve and prolong the life of the cath...

متن کامل

Fibrin sheath removal from central venous catheters: an internal snare manoeuvre.

BACKGROUND Dysfunction of haemodialysis catheters is most commonly due to a narrowing of the catheter lumen and/or formation of a fibrin sheath around the catheter tip. Reported methods for restoring patency of the catheter lumen include passage of a J-tipped guide wire, passage of a biopsy brush through the catheter, or infusion of a thrombolytic agent into the catheter. While these methods ar...

متن کامل

Dialysis catheter 'fibrin-sheath stripping': a cautionary tale!

Permcath@ (Quinton Instrument, Co. Seattle) was Introduction inserted and a primary arteriovenous fistula created in the operating theatre. Four weeks later the patient Dual-lumen cu ed catheters have assumed an increaspresented with a fever and subsequently Staphylococcus ingly important role for shortand long-term vascular aureus was cultured from his blood. The Permcath was access for the pr...

متن کامل

Fibrin sheaths in central venous port catheters: treatment with low-dose, single injection of urokinase on an outpatient basis

PURPOSE Evaluation of the efficacy of single-shot, low-dose urokinase administration for the treatment of port catheter-associated fibrin sheaths. METHODS Forty-six patients were retrospectively evaluated for 54 episodes of port catheter dysfunction. The presence of a fibrin sheath was detected by angiographic contrast examinations. On an outpatient basis, patients subsequently received throm...

متن کامل

Brachial Approach As an Alternative Technique of Fibrin Sheath Removal for Implanted Venous Access Devices

Implanted venous access device (IVAD) late dysfunction is commonly caused by fibrin sheath formation. The standard method of endovascular fibrin sheath removal is performed via the femoral vein. However, it is not always technically feasible and sometimes contraindicated. Moreover, approximately 4-6 h of bed rest is necessary after the procedure. In this article, we describe an alternative meth...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2012