Pulmonary vein antral isolation using an open irrigation ablation catheter for the treatment of atrial fibrillation: a randomized pilot study.

نویسندگان

  • Mohamed H Kanj
  • Oussama Wazni
  • Tamer Fahmy
  • Sergio Thal
  • Dimpi Patel
  • Claude Elay
  • Luigi Di Biase
  • Mauricio Arruda
  • Walid Saliba
  • Robert A Schweikert
  • Jennifer E Cummings
  • J David Burkhardt
  • David O Martin
  • Gemma Pelargonio
  • Antonio Dello Russo
  • Michela Casella
  • Pietro Santarelli
  • Domenico Potenza
  • Raffaele Fanelli
  • Raimondo Massaro
  • Giovanni Forleo
  • Andrea Natale
چکیده

OBJECTIVES We sought to test how catheter ablation using an open irrigation catheter (OIC) compares with standard catheters for pulmonary vein antrum isolation. BACKGROUND Open irrigation catheters have the advantage of delivering greater power without increasing the temperature of the catheter tip, which enables deeper and wider lesions without the formation of coagulum on catheters. METHODS Catheter ablation was performed using an 8-mm catheter (8MC) or an OIC. Patients were randomized to 3 groups: 8MC; OIC-1, OIC with a higher peak power (50 W); and OIC-2, OIC with lower peak power (35 W). RESULTS A total of 180 patients were randomized to the 3 treatment strategies. Isolation of pulmonary vein antra was achieved in all patients. The freedom from atrial fibrillation was significantly greater in the 8MC and OIC-1 groups compared with the OIC-2 group (78%, 82%, and 68%, respectively, p = 0.043). Fluoroscopy time was lower in OIC-1 compared with OIC-2 and 8MC (28 +/- 1 min, 53 +/- 2 min, and 46 +/- 2 min, respectively, p = 0.001). The mean left atrium instrumentation time was lower in the OIC-1 compared with the OIC-2 and 8MC groups (59 +/- 3 min, 90 +/- 5 min, and 88 +/- 4 min, respectively, p = 0.001). However, there was a greater incidence of "pops" in the OIC-1 (100%, 0%, 0%, p < 0.001) along with higher incidences of pericardial effusion (20%, 0%, 0%, p < 0.001) and gastrointestinal complaints (17% in OIC-1, 3% in 8MC, and 5% in OIC-2, p = 0.031). CONCLUSIONS Although there was a decrease in fluoroscopy and left atrium instrumentation time with the use of OIC at higher power, this setting was associated with increased cardiovascular and gastrointestinal complications.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Single Center Experience with a Closed-Loop Irrigated Ablation Catheter for the Treatment of Human Paroxysmal and Persistent Atrial Fibrillation

Open-irrigated electrode catheters are commonly used for ablation for atrial fibrillation (AF). There are scant data available for closed-loop irrigation for ablation for AF. Proponents of open-irrigated catheters state greater patient safety from char formation, less impedance rises, and greater lesion depth. There are fundamental biophysical differences in the biophysics of the energy deliver...

متن کامل

Effect of Oral Anticoagulant Therapy on Coagulation Activity and Inflammatory Markers in Patients with Atrial Fibrillation Undergoing Ablation: A Randomized Comparison between Dabigatran and Warfarin

  Atrial fibrillation (AF) is associated with inflammatory and hypercoagulability state. Previous studies evaluated the safety and efficacy of dabigatran and warfarin in prevention of thrombothic complications. This study was intended to assess the influence of these drugs on hemostatic and inflammatory markers among patient underwent pulmonary vein ablation.   A total of 100 patients with AF w...

متن کامل

Arrhythmias: drugs and devices.

Arrhythmias of supraventricular and ventricular origin are an important health issue due to mortality, morbidity, and the impaired quality of life of patients with which it is associated, but are also economically relevant due to the physician’s visits, hospitalizations, and interventions they lead to. For certain arrhythmias such as bradycardia of different origin and ventricular tachycardia a...

متن کامل

Effect of Oral Anticoagulant Therapy on Coagulation Activity and Inflammatory Markers in Patients with Atrial Fibrillation Undergoing Ablation: A Randomized Comparison between Dabigatran and Warfarin

  Atrial fibrillation (AF) is associated with inflammatory and hypercoagulability state. Previous studies evaluated the safety and efficacy of dabigatran and warfarin in prevention of thrombothic complications. This study was intended to assess the influence of these drugs on hemostatic and inflammatory markers among patient underwent pulmonary vein ablation.   A total of 100 patients with AF w...

متن کامل

Multielectrode Pulmonary Vein Isolation Versus Single Tip Wide Area Catheter Ablation for Paroxysmal Atrial Fibrillation

Catheter ablation (CA) is an accepted treatment option for symptomatic atrial fibrillation (AF). Historically, CA has been performed with a point-by-point technique, using standard focal radiofrequency catheters and 3-dimensional (3D) navigation systems, with a procedural end point of electric pulmonary vein isolation (PVI). However, this CA approach requires advanced training and operator skil...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of the American College of Cardiology

دوره 49 15  شماره 

صفحات  -

تاریخ انتشار 2007