Five-year clinical outcome and patency rate of device-dependent venous grafts after clampless OPCAB with PAS-port automated proximal anastomosis: the PAPA Study.

نویسندگان

  • Carlo Bassano
  • Emanuele Bovio
  • Massimiliano Sperandio
  • Floriano Uva
  • Andrea Farinaccio
  • Paolo Prati
  • Luigi Chiariello
چکیده

OBJECTIVE To evaluate long-term clinical performance and angiographic patency of automated proximal venous anastomoses following clampless coronary artery bypass (C-CAB). METHODS Observational study in patients submitted for isolated C-CAB and at least one proximal aortosaphenous anastomosis performed with an automated connector (Cardica PAS-Port) including 152 consecutive patients (165 devices and 199 device-dependent distal anastomoses), with LVEF > 30% and saphenous vein diameter of 4-6 mm. Clinical follow-up was 96% complete (4101/4269 pt-months). Graft patency rate was assessed with 64-slice CT-scan or coronary angiography. Freedom from major adverse cardiac and cerebrovascular events (MACCE) was reported as actuarial probability with 95% confidence limits and venous graft patency as actual rate at every year interval. RESULTS Early operative mortality was 1.9%; incidence of neurologic injury was zero. Freedom from MACCE was 92.7 ± 2.1 at one year and 85.2 ± 4.8 at five years. The actual patency rate of device-dependent venous grafts was 90%, 85%, 84%, 84%, and 93% for one-, two-, three-, four-, and five-year-old grafts, respectively. CONCLUSIONS The device is a well-performing system for proximal anastomoses. The incidence of neurologic complications seems to be reduced with this clampless approach. The high patency rate is stable over time.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Clampless off-pump coronary artery bypass prevents early post-operative neurologic injury compared to on-pump: a propensity score matched analysis

Background Unaortic coronary artery bypass with all arterial grafts proved to prevent early neurologic injury. Various devices for clampless proximal anastomosis of venous grafts were associated with an increased risk of microembolism. The Cardica PAS-Port is a new, fully automated device that might be able to obtain safe proximal aorto-venous connection without an increased risk of microor mac...

متن کامل

Twelve-month patency with the PAS-port proximal connector device: a single center prospective randomized trial. Invited commentary.

BACKGROUND The PAS-Port (Cardica Inc, Redwood City, CA) is an automated system that allows for the clampless anastomosis of vein grafts to the aorta. The intent of this study was to prospectively compare one year graft patency of this system with conventional hand-sewn anastomoses in a prospective, randomized trial. METHODS A total of 99 patients undergoing elective off-pump coronary bypass s...

متن کامل

Six-month angiographic follow-up of the PAS-Port II clinical trial.

BACKGROUND The PAS-Port device (Cardica, Redwood City, CA) allows the rapid deployment of a clampless proximal anastomosis between a vein graft and the aorta. METHODS Fifty-four patients awaiting elective coronary artery bypass graft surgery were enrolled. Outcome variables were intraoperative device performance, early and 6- month angiographic graft patency, and 12-month clinical follow-up. ...

متن کامل

Morphological Evaluation of Proximal Anastomosis by PAS-Port(®) System in Patients with Long-Term Patent Grafts.

PURPOSE We examined grafts employing for morphological analysis of early and long-term results on proximal anastomosis with the PAS-Port(®). METHODS One hundred and four patients treated by OPCAB with PAS-Port(®) were performed postoperative MDCT. Morphological evaluation of the proximal anastomotic region was classified into three groups (A; graft was anastomosed almost perpendicularly to th...

متن کامل

Evaluation of the PAS-Port Proximal Anastomosis System in coronary artery bypass surgery (the EPIC trial).

OBJECTIVE During coronary surgery, proximal vein graft anastomoses have been performed by using an aortic partial occlusion clamp to allow for a hand-sewn anastomosis. The purpose of this multicenter, prospective, randomized trial was to evaluate the efficacy of the PAS-Port device (Cardica, Inc, Redwood City, Calif), which allows an automated proximal anastomosis to be performed without aortic...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of cardiac surgery

دوره 29 3  شماره 

صفحات  -

تاریخ انتشار 2014