The use of a coin shaped implant for direct in situ measurement of attachment strength for osseointegrating biomaterial surfaces.

نویسندگان

  • Hans Jacob Rønold
  • Jan Eirik Ellingsen
چکیده

Most animal models currently used to study the retention of implants in bone are influenced by shear forces introduced during the retention test. This is mainly due to the implant design, which most often are cylindrical, conical or threaded. In these models interlocking between bone and implant surface will increase the effect of genuine bone bonding and thus give a false positive outcome. The purpose of the present study was to establish a model for testing functional attachment of implants in situ, with minimal influence of interlocking and shear forces. The model involves the use of flat coin shaped implant placed onto the cortical bone of rabbit tibia without mechanical fixation to the bone. The implant is passively retained on the cortical bone by a titanium band retainer. During the healing period, the contact between the coin shaped implants and the bone is restricted to the flat test surfaces. To prevent interlocking effects from lateral bone attachments a polytetrafluoroethylene (PTFE) cap covering the vertical and the upper faces of the implants were used. The tensile test was performed with a gradual, calibrated pull, perpendicular to the bone-implant interface. This pullout model makes it possible to study the kinetics and strength of bone bonding with negligible influence of shear forces or mechanical interlocking.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of micro-roughness produced by TiO2 blasting--tensile testing of bone attachment by using coin-shaped implants.

The aim of the present study was to examine bone response to micro-rough titanium implants. Forty coin-shaped implants were divided into eight groups according to their surface roughness. The first group had electropolished surfaces. The surfaces of implant groups 2-8 were blasted with TiO2 particles with incremental grain sizes ranging from 7.5-12.5 to 270-330 microns. Five implants from each ...

متن کامل

Stress Distribution in Natural Tooth and Implant Supported Removable Partial Denture with Different Attachment Types: A Photoelastic Analysis

Background and Aim: Different attachment designs have been developed to connect implant to natural teeth in partial dentures; however, adequate studies have not been performed to determine stress distribution patterns in these designs. The present study aimed to assess stress distribution patterns in natural tooth and implant supported removable partial denture with different attachment designs...

متن کامل

Target Strength Measurement of Myctophid (Benthosema pterotum) in the Oman Sea

Through a series of research cruises conducted by the R/V Ferdows-1 to assess Myctophid resources, using a SIMRAD EK500 scientific echosounder carried out the Target Strength measurement of Benthosema pterotum in the Oman Sea. The principal method was based on a direct in-situ split-beam measurement technique, which is implemented in the EK500 system. For real-time observations ,the target st...

متن کامل

Mode III cleavage of a coin-shaped titanium implant in bone: effect of friction and crack propagation.

Endosseous cementless implants are widely used in orthopaedic, maxillofacial and oral surgery. However, failures are still observed and remain difficult to anticipate as remodelling phenomena at the bone-implant interface are poorly understood. The assessment of the biomechanical strength of the bone-implant interface may improve the understanding of the osseointegration process. An experimenta...

متن کامل

Effect of UV-Photofunctionalization on Bioactivity of Titanium to Promote Human Mesenchymal Stem Cells

Background and Aim: The present study introduces photofunctionalization as a technique for tackling biological aging and increasing the bioactivity of titanium. This in-vitro study evaluated the effects of ultraviolet (UVC) light treatment of titanium surfaces with different time-related changes on the behavior and function of human mesenchymal stem cells (MSCs). Materials and Methods: MSCs we...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Biomaterials

دوره 23 10  شماره 

صفحات  -

تاریخ انتشار 2002