DPOAE input/output (I/O) functions were measured at 7f2 frequencies (1 to 8 kHz; f2/f1 = 1.22) over a range of levels (-5 to 95 dB SPL) in normal-hearing and hearing-impaired human ears. L1-L2 was level dependent in order to produce the largest 2f1-f2 responses in normal ears. System distortion was determined by collecting DP data in six different acoustic cavities. These data were used to derive a multiple linear regression model to predict system distortion levels. The model was tested on cochlear-implant users and used to estimate system distortion in all other ears. At most but not all f2's, measurements in cochlear implant ears were consistent with model predictions. At all f2 frequencies, the ears with normal auditory thresholds produced I/O functions characterized by compressive nonlinear regions at moderate levels, with more rapid growth at low and high stimulus levels. As auditory threshold increased, DPOAE threshold increased, accompanied by DPOAE amplitude reductions, notably over the range of levels where normal ears showed compression. The slope of the I/O function was steeper in impaired ears. The data from normal-hearing ears resembled direct measurements of basilar membrane displacement in lower animals. Data from ears with hearing loss showed that the compressive region was affected by cochlear damage; however, responses at high levels of stimulation resembled those observed in normal ears.
منابع مشابه
Evaluation of Distortion and Transient Evoked Otoacoustic Emission in Tinnitus Patients with Normal Hearing
Introduction: Tinnitus is a perception of sound without external source. The exact etiology of tinnitus is not fully understood, although some researchers believe that the condition usually starts in the cochlea. The aim of this study was to determine the potential contribution of outer hair cell dysfunction to chronic tinnitus, by application of Distortion-Product Evoked Otoacoustic Emission (...
متن کاملInput-output functions of the nonlinear-distortion component of distortion-product otoacoustic emissions in normal and hearing-impaired human earsa)
Distortion-product otoacoustic emissions (DPOAEs) arise in the cochlea in response to two tones with frequencies f1 and f2 and mainly consist of two components, a nonlinear-distortion and a coherent-reflection component. Wave interference between these components limits the accuracy of DPOAEs when evaluating the function of the cochlea with conventional continuous stimulus tones. Here, DPOAE co...
متن کاملDistortion product otoacoustic emission suppression tuning curves in normal-hearing and hearing-impaired human ears.
Distortion product otoacoustic emission (DPOAE) suppression measurements were made in 20 subjects with normal hearing and 21 subjects with mild-to-moderate hearing loss. The probe consisted of two primary tones (f2, f1), with f2 held constant at 4 kHz and f2/f1 = 1.22. Primary levels (L1, L2) were set according to the equation L1 = 0.4 L2 + 39 dB [Kummer et al., J. Acoust. Soc. Am. 103, 3431-34...
متن کاملOtoacoustic emissions from normal-hearing and hearing-impaired subjects: distortion product responses.
Distortion product otoacoustic emissions (DPOAE) were measured in normal-hearing and hearing-impaired human subjects. Analyses based on decision theory were used to evaluate DPOAE test performance. Specifically, relative operating characteristic (ROC) curves were constructed and the areas under these curves were used to estimate the extent to which normal and impaired ears could be correctly id...
متن کاملThe effects of high-frequency hearing loss on low-frequency components of the click-evoked otoacoustic emission.
Click-evoked otoacoustic emission (CEOAE) input/output (I/O) functions were measured in ears with normal hearing and in ears with sensorineural hearing loss above 2000 Hz. The low- to midfrequency CEOAEs obtained from the ears with high-frequency hearing loss were significantly reduced in level compared to the CEOAEs obtained from the ears with normal hearing even though there were no significa...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of the Acoustical Society of America
دوره 110 6 شماره
صفحات -
تاریخ انتشار 2001