Is the PASCAL-Tonometer suitable for measuring intraocular pressure in clinical routine? Long- and short-term reproducibility of dynamic contour tonometry.
نویسندگان
چکیده
PURPOSE Dynamic contour tonometry (DCT) is a new technique to measure intraocular pressure (IOP). In several studies no correlation between IOP values and corneal thickness was shown with DCT. This is in contrast to the gold standard, Goldmann applanation tonometry (GAT). The authors evaluated the reproducibility (RP) of DCT compared to GAT, a prerequisite for its introduction into clinical routine. METHODS IOP was measured with both DCT and GAT in 50 subjects with normal cornea. To evaluate the short-term RP, four DCT and four GAT measurements were performed at day 1 in a randomized order. Long-term RP was determined by one additional measurement per method at day 2, 5, and 8. RESULTS The short-term RP was defined as the mean value of all standard deviations (SD) of the individual measurements at day 1. Short-term RP was 1.1 mmHg for GAT and 1.2 mmHg for DCT. For long-term reproducibility, mean SD was 1.2 mmHg for GAT and 1.5 mmHg for DCT. Bland-Altman revealed a good agreement of the two methods. However, mean DCT values were on average 0.8+/-1.1 mmHg higher than GAT values. A significant correlation was observed between GAT and CCT (r2=0.15, p=0.006), but not between DCT and CCT (r2=0.032, p=0.21). CONCLUSIONS Short- and long-term reproducibility of DCT is comparable to that of GAT. GAT is more affected by CCT than DCT, measuring higher IOPs in eyes with higher central corneal thickness.
منابع مشابه
History and Clinical Validation of the PASCAL Dynamic Contour Tonometer
Will Goldmann applanation tonometry, the age old “gold standard” of intraocular pressure (IOP) measurement, be replaced with newer methods? One of the newer methods of IOP measurement is PASCAL dynamic contour tonometry (DCT). A review of the history, scientific principles, clinical validation and clinical utility of DCT is presented.
متن کامل[Comparison of dynamic contour tonometry (Pascal) with pneumotonometry and Goldmann tonometry].
PURPOSE To compare the intraocular pressure measurements as defined by the Pascal tonometer, the Goldmann tonometer and the pneumotonometer. METHODS This was an observational clinical study, which included two hundred and five randomly selected subjects recruited from the Ophthalmology Department. The intraocular pressure measurements were performed with each tonometry technique in a randomiz...
متن کاملDynamic contour tonometry and goldman applanation tonometry in eyes with keratoconus.
PURPOSE The dynamic contour tonometer is a nonapplanation contact tonometer designed to be largely independent of the structural properties of the cornea. Theoretically, it may measure intraocular pressure most accurately in abnormally thinner corneas. This study compares intraocular pressure measurements by dynamic contour tonometry with Goldman applanation tonometry in eyes with normal cornea...
متن کاملReview on Dynamic Contour Tonometry and Ocular Pulse Amplitude.
Intraocular pressure (IOP) measurement is the cornerstone of the management of glaucoma patients. The gold standard for assessing IOP is Goldmann applanation tonometry (GAT). Recently, the dynamic contour tonometer (DCT) has become available. While both devices provide reliable IOP measurements, the results are not interchangeable. DCT has the advantage of measuring an additional parameter: ocu...
متن کاملThe relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry.
AIMS To establish the effects of central corneal thickness (CCT) on intraocular pressure (IOP) measured with a prototype Pascal dynamic contour tonometer (DCT), to evaluate the effect of CCT and age on the agreement between IOP measured with the Pascal DCT and Goldmann applanation tonometer (GAT), and to compare the interobserver and intraobserver variation of the DCT with the GAT. METHODS GA...