Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use?

نویسندگان

  • John N Amoore
  • David Dewar
  • Kathleen Gough
  • Paul L Padfield
چکیده

OBJECTIVE To assess the measurement consistency of SpaceLabs ambulatory recorders (Spacelabs, Washington, USA) that are in regular use. METHODS A total of 14 SpaceLabs 90207 and one 90217 ambulatory recorders were tested for measurement consistency using the Dynatech CuffLink (Dynatech, Nevada, USA), a commercially available non-invasive blood pressure (NIBP) simulator. The NIBP recorders were tested at a range of pressures with 20 repeated determinations at a simulated 120/80 mmHg and five repeated determinations at simulated pressures of 80/50, 100/80, 150/100, 200/165 and 250/195 mmHg. Tests were carried out in 1998, 2002 and late 2003 or early 2004. CONCLUSIONS All 15 SpaceLabs recorders measured consistently over the 6 years with 89.5% of the differences in average pressures, recorded by any particular device at each recorded pressure, less than 2 mmHg between successive test episodes. The maximum difference was 4.5 mmHg and 60.1% of the differences were less than 1 mmHg. The measurements for all devices were within the tolerances specified by the supplier for the device when tested with the simulator. Maintenance records also show that most devices required breakdown maintenance less than once every 3 years. The results show that the SpaceLabs devices maintain measurement consistency in the demanding conditions of ambulatory pressure recording over several years.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the SpaceLabs 90202 non-invasive ambulatory recorder according to the AAMI Standard and BHS criteria.

The SpaceLabs 90202, a non-invasive ambulatory blood pressure recorder for the measurement of 24 hr blood pressure, was assessed according to the standard of the Association for the Advancement of Medical Instrumentation (AAMI) and the grading criteria of the British Hypertension Society (BHS) protocol were applied to the results. Two observers measured BP simultaneously in the same arm with th...

متن کامل

Ambulatory blood pressure measurement in the diagnosis and management of hypertension.

Before the diagnostic potential of 24-hour non-invasive BP measurement can be assessed, the accuracy of ambulatory recorders must be established, and normal reference values determined. The accuracy criteria of four ambulatory BP measuring systems (the SpaceLabs 90207, the Novecor DIASYS 200, the Takeda TM-2420 and the Del Mar Avionics Pressurometer IV) have been assessed according to the Briti...

متن کامل

Using Hawksley random zero sphygmomanometer as a gold standard may result in misleading conclusions.

We combined a database of paired blood pressure measurements taken using the Hawksley random-zero sphygmonanometer and a standard mercury sphygmomanometer and a database of paired measurements made on a SpaceLabs 90202 ambulatory blood pressure recorder and standard sphygmomanometer to determine how the SpaceLabs 90202 would have fared if it had been assessed against the Hawksley random-zero sp...

متن کامل

Bias and variability in blood pressure measurement with ambulatory recorders.

This study sought to determine whether patient characteristics such as age, sex, blood pressure, and pulse pressure differently affect the accuracy of an oscillometric (SpaceLabs 90207) and a microphonic (TM2420 version 7) blood pressure monitor. Blood pressure recorded by two oscillometric and two microphonic ambulatory monitors was compared with simultaneous readings by two pairs of trained, ...

متن کامل

Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.

This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial dias...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Blood pressure monitoring

دوره 10 1  شماره 

صفحات  -

تاریخ انتشار 2005