New roller pump disposable provides safety and simplifies occlusion setting.

نویسندگان

  • K Lee-Sensiba
  • N Azzaretto
  • C Carolina
  • N DiCarmine
  • D Hymowitz
  • S Kay
  • K Kooker
  • M Salogub
  • E Wong
  • Y Tamari
چکیده

A new disposable insert for the arterial roller pump, the Better-Header, provides safety and functionality beyond what standard tubing provides. It automatically limits pump outlet pressure to a level determined by the user and provides a self-contained, simple means to set pump occlusion. The Better-Header consists of a Starling-like pressure relief valve connected across standard header tubing. As long as arterial line pressure at the pump outlet remains below a set limit, the valve is closed. If line pressure approaches the pressure limit, the valve opens, preventing overpressurization by shunting blood from pump outlet to inlet. The Better-Header can also be used to set occlusion by the "dynamic method" to obtain nonocclusive settings. The Better-Header was evaluated in the lab for its pressure-flow characteristics. Even when the arterial line was completely clamped at a pump flow of 7 L/min, line pressure was limited to a safe level and all circuit connections were preserved. The Better-Header has been used successfully at North Shore University Hospital in over 500 clinical cases covering a wide range of patients and procedures. In several instances, the user was alerted to high pressure situations by fluid flow through the valve and by an audible alarm, allowing rapid correction of the source of pressure. Compared to the standard setup, the Better-Header maintains outlet pressure within safe, user-settable limits, and permits consistent, nonocclusive settings with predictable retrograde flow.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Comparison of Static Occlusion Setting Methods : Fluid Drop Rate and Pressure Drop

Tubing circuits primed with water were constructed on two roller pumps to characterize and compare two static methods of setting the occlusion on roller pumps (fluid drop rate and pressure drop rate). Twelve separate experiments were performed in which PVC boot tubing diameters (1/2 in, 3/8 in and 1/4 in), roller position (each roller alone on a two roller pump) and roller occlusion (total occl...

متن کامل

An under-occluded roller pump is less hemolytic than a centrifugal pump.

The purpose of this study is to measure and compare the hemolysis produced by roller pumps with varied occlusion settings and a centrifugal pump. The null hypothesis is that there is no difference in the Index of Hemolysis (IH = gm Hb/100 L pumped) produced by a roller pump (RP) at four different occlusion settings and a centrifugal pump (CP) at the same blood flow rate (4.5 L/min) and afterloa...

متن کامل

Significant safety advantages gained with an improved pressure-regulated blood pump.

A prototype of a non-occlusive pressure-regulated blood pump (M-pump) was evaluated in-vitro for safety in a comparative study with the roller and centrifugal pumps. The M-pump consists of collapsible tubing of unique design wrapped under tension around a rotor without a stator. The prototype M-pumps were tested in vitro to evaluate performance with respect to flow/ pressure characteristics, he...

متن کامل

Non-occlusive condition with the Better-Header roller pump: impacts of flow dynamics and hemolysis.

PURPOSE Impacts on hemolysis and backflow using a non-occlusive setting with the Better-Header (BH) roller pump were investigated. METHODS AND RESULTS Pump flow of a non-occlusion setting was measured with a pump speed of 3 L/min and 5 L/min against various after-loads. With the non-occlusive setting (BH-NO350), backflow was less than 10% if the pump pressure head was <300 mmHg. When the outl...

متن کامل

Laboratory evaluation of the limitations of positive pressure safety valves on hard-shell venous reservoirs.

Vacuum-assisted venous drainage (VAVD) is a technique used to increase venous return during cardiopulmonary bypass (CPB). However, VAVD has created some new safety concerns. One potential problem is the pressurization of the venous reservoir in the event of vacuum failure. To prevent this overpressurization, a positive pressure release valve (PPRV) is placed on the venous reservoir. The purpose...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 29 1  شماره 

صفحات  -

تاریخ انتشار 1997