A simple method of measuring the wear of explanted acetabular component inserts

نویسندگان

  • L. Krakow
  • A. Klockow
  • E. Roehner
  • S. Brodt
  • H. Eijer
  • J. Bossert
  • G. Matziolis
چکیده

AIMS The determination of the volumetric polyethylene wear on explanted material requires complicated equipment, which is not available in many research institutions. Our aim in this study was to present and validate a method that only requires a set of polyetheretherketone balls and a laboratory balance to determine wear. METHODS The insert to be measured was placed on a balance, and a ball of the appropriate diameter was inserted. The cavity remaining between the ball and insert caused by wear was filled with contrast medium and the weight of the contrast medium was recorded. The volume was calculated from the known density of the liquid. The precision, inter- and intraobserver reliability, were determined by four investigators on four days using nine inserts with specified wear (0.094 ml to 1.626 ml), and the intra-class correlation coefficient was calculated. The feasibility of using this method in routine clinical practice and the time required for measurement were tested on 84 explanted inserts by one investigator. RESULTS In order to get the mean for all investigators and determinations, the deviation between the measured and specified wear was -0.08 ml (sd 0.12; -0.21 to 0.11). The interobserver reliability was 0.989 ml (95% confidence interval (CI) 0.964 to 0.997) and the intraobserver reliability was 0.941 for observer 1 (95% CI 0.846 to 0.985), 0.983 for observer 2 (95% CI 0.956 to 0.995), 0.939 for observer 3 (95% CI 0.855 to 0.984), and 0.934 for observer 4 (95% CI 0.790 to 0.984). The mean time required to examine the samples was two minutes (sd 2; 1 to 5). CONCLUSION The method presented here was shown to be sufficiently precise for many settings and is a cost-effective and quick method of determining the volumetric wear of explanted acetabular components. However, the measurement of wear for scientific purposes will probably continue to involve more accurate and dedicated laboratory equipment.Cite this article: Bone Joint Res 2017;6:530-534.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Method for the location of primary wear scars from retrieved metal on metal hip replacements

BACKGROUND Retrieved metal-on-metal acetabular cups are valuable resources in investigating the wear behaviour of failed hip implants, but adequate methods to do so are lacking. To further contribute to addressing this issue, we developed a method to detect the in vivo location of the primary wear scar of an explanted cup. METHODS We proposed a new method in which thirteen patients with faile...

متن کامل

Measurement of femoral head penetration in polyethylene using a 3-dimensional CT-scan technique

BACKGROUND Current techniques for measuring in vivo polyethylene wear suffer from a range of problems, resulting in an unacceptable lack of repeatability and/or insufficient accuracy when they are used to measure the low wear rates associated with new, highly crosslinked polyethylene. We describe an improved CT method for measurement of 3D femoral head penetration in PE acetabular cups that has...

متن کامل

Ceramic-on-metal bearings in total hip replacement: whole blood metal ion levels and analysis of retrieved components.

This study reports on ceramic-on-metal (CoM) bearings in total hip replacement. Whole blood metal ion levels were measured. The median increase in chromium and cobalt at 12 months was 0.08 microg/1 and 0.22 microg/1, respectively, in CoM bearings. Comparable values for metal-on-metal (MoM) were 0.48 microg/1 and 0.32 microg/1. The chromium levels were significantly lower in CoM than in MoM bear...

متن کامل

A comparison of explanted Articular Surface Replacement and Birmingham Hip Resurfacing components.

The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the c...

متن کامل

In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty

BACKGROUND Determination of the amount of wear in a polyethylene liner following total hip arthroplasty (THA) is important for both the clinical care of individual patients and the development of new types of liners. PATIENTS AND METHODS We measured in vivo wear of the polyethylene liner using computed tomography (CT) (obtained in the course of regular clinical care) and compared it to coordi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 6  شماره 

صفحات  -

تاریخ انتشار 2017