Comparison of ocular biometry measurements by applanation and immersion A-scan techniques

نویسندگان

  • Dupe S. Ademola-Popoola
  • Donald A. Nzeh
  • Sadiat E. Saka
  • Lateefat B. Olokoba
  • Tokunbo S. Obajolowo
چکیده

PURPOSE The study compared ocular biometry values using applanation and immersion techniques to determine the most applicable method for our tertiary training centre where personnel with different levels of experience and expertise in biometry take measurements used in calculation of required intraocular lens before cataract surgery. METHODS The study was a prospective cross-sectional comparative study of different techniques of ocular biometry from diagnostic equipment (biometry probe 10 MHz Sonomed(®) A-scan (PACSCAN 300A, USA). Measurement variables were obtained among children and adults undergoing cataract surgery. Scleral (Prager) shell was used for the immersion technique followed by the contact technique by the same examiner. RESULTS The biometry values of 92 eyes of 92 adult were taken. Their ages ranged from 18 to 95 years with a mean of 64.7 (SD ± 12.9) years. There were 55 (59.8%) males and 37 (40.2%) females, with a male to female ratio of 1.5:1. Average axial length (22.0-24.4 mm) eyes were the most common eyes measured in 75 (81.5%) of the cases. The means of the axial lengths biometry values with immersion and contact technique were 23.66(±1.36) and 23.46 mm (±1.46); the axial length differences was 0.2 ± 0.26 mm (range 0.0-0.94 mm) and statistically significant (95% CI of the Difference 0.15 to 0.26, p = 0.000). The Standard deviation SD (mm) of Individual Eye Axial Length showed a mean of 0.03 ± 0.04 (0.0-0.3) mm for immersion and for contact technique 0.14 ± 0.12(0.0-0.6)mm. CONCLUSION There was a significant difference in ocular biometry measurement with the contact and immersion ultrasound techniques. The immersion technique had better repeatability, thus it is ideal in a training hospital setting in a typical in sub-Saharan Africa who have limited resources to employ a dedicated person to do biometry; and where the different operators of A-scan machines have different levels of experience and expertise.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Preoperative and Postoperative Ocular Biometry in Eyes with Phakic Intraocular Lens Implantations

PURPOSE To compare preoperative and postoperative ocular biometry in patients with iris-fixated phakic intraocular lens (pIOLs): Artisan and Artiflex. MATERIALS AND METHODS This study included 40 eyes with Artisan and 36 eyes with Artiflex pIOL implants. Anterior chamber depth (ACD) and axial length (AL) were measured by applanation ultrasonography (A-scan) and partial coherence interferometr...

متن کامل

The Short-Term Effects of Exercise on Intraocular Pressure, Choroidal Thickness and Axial Length

PURPOSE To explore ocular changes in healthy people after exercise. METHODS Twenty five volunteers underwent exercise for 15 minutes on a treadmill. Measurements of choroidal thickness, intraocular pressure (IOP), ocular biometry, and blood pressure were taken before and after exercise. Enhanced Depth Imaging optical coherence tomography (EDI-OCT) was used to measure choroidal thickness at th...

متن کامل

A new non-contact optical device for ocular biometry.

BACKGROUND A new commercially available device (IOLMaster, Zeiss Instruments) provides high resolution non-contact measurements of axial length (using partial coherent interferometry), anterior chamber depth, and corneal radius (using image analysis). The study evaluates the validity and repeatability of these measurements and compares the findings with those obtained from instrumentation curre...

متن کامل

CLINICAL SCIENCE A new non-contact optical device for ocular biometry

Background: A new commercially available device (IOLMaster, Zeiss Instruments) provides high resolution non-contact measurements of axial length (using partial coherent interferometry), anterior chamber depth, and corneal radius (using image analysis). The study evaluates the validity and repeatability of these measurements and compares the findings with those obtained from instrumentation curr...

متن کامل

Accuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan

PURPOSE To compare the refractive results of cataract surgery measured by applanation ultrasound and the new partial coherence interferometer, AL-scan. METHODS Medical records of 76 patients and 104 eyes who underwent cataract surgery from January 2013 to June 2013 were retrospectively reviewed. Biometries were measured using ultrasound and AL-scan and intraocular lens power was calculated us...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 27  شماره 

صفحات  -

تاریخ انتشار 2015