Target and peripheral dose during patient repositioning with the Gamma Knife automatic positioning system (APS) device
نویسندگان
چکیده
The GammaPlan treatment planning system does not account for the leakage and scatter dose during APS repositioning. In this study, the dose delivered to the target site and its periphery from the defocus stage and intershot couch transit (couch motion from the focus to defocus position and back) associated with APS repositioning are measured for the Gamma Knife model 4C. A stereotactic head-frame was attached to a Leksell 16 cm diameter spherical phantom with a calibrated ion chamber at its center. Using a fiducial box, CT images of the phantom were acquired and registered in the GammaPlan treatment planning system to determine the coordinates of the target (center of the phantom). An absorbed dose of 10 Gy to the 50% isodose line was prescribed to the target site for all measurements. Plans were generated for the 8, 14 and 18 mm collimator helmets to determine the relationship of measured dose to the number of repositions of the APS system and to the helmet size. The target coordinate was identical throughout entire study and there was no movement of the APS between various shots. This allowed for measurement of intershot transit dose at the target site and its periphery. The couch was paused in the defocus position, allowing defocus dose measurements at the intracranial target and periphery. Measured dose increases with frequency of repositioning and with helmet collimator size. During couch transit, the target receives more dose than peripheral regions; however, in the defocus position, the greatest dose is superior to the target site. The automatic positioning system for the Leksell Gamma Knife model 4C results in an additional dose of up to 3.87 +/- 0.07%, 4.97 +/- 0.04%, and 5.71 +/- 0.07% to the target site; its periphery receives additional dose that varies depending on its position relative to the target. There is also dose contribution to the patient in the defocus position, where the APS repositions the patient from one treatment coordinate to another. This may be important for treatment areas around critical structures within the brain. Further characterization of the defocus and transit exposures and development of a dose calculation algorithm to account for these doses would improve the accuracy of the delivered plan.
منابع مشابه
Target and peripheral dose from radiation sector motions accompanying couch repositioning of patient coordinates with the Gamma Knife® Perfexion™
BACKGROUND The GammaPlan(™) treatment planning system (TPS) does not fully account for shutter dose when multiple shots are required to deliver a patient's treatment. The unaccounted exposures to the target site and its periphery are measured in this study. The collected data are compared to a similar effect from the Gamma Knife(®) model 4C. MATERIALS AND METHODS.: A stereotactic head frame wa...
متن کاملEvaluation of Organ Specific peripheral dose for Gamma knife 4C based on Monte Carlo
Introduction: Stereotactic Gamma Knife radiosurgery has been widely used for treating brain tumors. The scattered radiation outside of treatment field (peripheral dose) can induce the secondary cancer to specific organ. This paper investigated the absorbed dose to eyes, thyroid, heart, lung, breast and colon using a Monte Carlo technique for Mird phantom. We also study the ef...
متن کاملGamma Knife Surgery for Essential Trigeminal Neuralgia: New Treatment Strategy with Robotized Micro-Radiosurgery
Gamma knife surgery (GKS) is a minimally invasive treatment for brain diseases, and is currently used for functional disorders. Moreover, a revised version of Gamma knife system, “Model C-APS (automatic positioning system)” started to be installed in Japan since 2001. This latest model enables us to adjust coordinates by every 0.1 mm, and treat the patient automatically. We have completed a ret...
متن کاملRADIATION PROTECTION OF PATIENT DURING STEREOTACTIC RADIOSURGERY. Part I: Single isocenter phantom measurements
Radiation dose to organs outside the radiotherapy treatment field can be significant and clinical interest. Therefore, in order to estimate peripheral doses (PD) during the stereotactic radiosurgery treatment with Leksell Gamma Knife for single isocenter treatment technique in particular as a part of the method can be described the all plan treatment in general. Knowledge of PD can be used effe...
متن کاملEvaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife
Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 11 شماره
صفحات -