Effects of converting bilateral cochlear implant subjects to a strategy with increased rate and number of channels.
نویسندگان
چکیده
OBJECTIVES Three different Advanced Bionics processing strategies were evaluated: 1) 8-channel, 813 pulses per second (pps), Continuous Interleaved Sampling (CIS); 2) 16-channel, 5,100 pps, HiResolution Paired (HiRes P); and 3) 16-channel, 2,900 pps, HiResolution Sequential (HiRes S). METHODS Seven adult bilateral Clarion CII cochlear implant recipients who had been using a CIS processing strategy for at least 18 months participated in this study. Sentence recognition in multitalker babble from the front was collected on subjects using their CIS strategy and after subjects were programmed for the first time with HiRes P and HiRes S strategies. An ABAB design was implemented for 1 month whereby subjects used each HiResolution strategy every other day. Sentence recognition testing was repeated at the 1- and 3-month post-HiResolution test intervals. RESULTS Comparisons between the CIS and HiResolution strategies showed immediate improvements for 5 subjects in favor of the HiResolution strategies. After 1 month of alternating between the HiRes P and HiRes S strategies, remarkably, 2 subjects showed improvements of 60%, 2 subjects showed improvements of 40%, and 2 subjects showed improvements of 30% over the CIS strategy that they had previously used for at least 18 months. The results after 3 months of use were consistent with those obtained at 1 month. CONCLUSIONS The HiRes S and HiRes P strategies resulted in dramatic improvements in speech perception in noise for a subset of subjects who had been using the CIS strategy bilaterally. This finding demonstrates that these subjects were able to tolerate a more difficult signal-to-noise ratio. Further work is needed to determine the independent effects of rate versus number of channels.
منابع مشابه
COCHLEAR IMPLANTATION IN IRAN
Cochlear implantation has become an increasingly common procedure in the rehabilitation of selected cases of profound deafness.' Patients should have profound total bilateral sensorineural hearing loss. Sound is transformed into small electric currents which stimulate the auditory nerves in the cochlea and generate the hearing sensation. The nucleus cochlear implant is the result of more t...
متن کاملEvaluation of Saccular Function Pre-Post Cochlear Implant Surgery Using VEMPs
Objectives: The aim of present study was evaluation of saccule function in cochlear implant candidates with severe to profound sensory neural hearing loss Before and after cochlear implantation Methods: In this study 35 cochlear implant (CI) candidates with bilateral severe to profound sensory neural hearing loss before and about 30 days after cochlear implant and 20 normal-hearing cases as ...
متن کاملUse of Intrathecal Fluorescein in Recurrent Meningitis after Cochlear Implantation-A Case Report
Introduction: Congenital anomalies of the cochlea and labyrinth can be associated with meningitis and varying degrees of hearing loss or deafness. Despite antibiotics, meningitis remains a life threatening complication. Case Report: We report a case of recurrent meningitis following episodes of otitis media in a cochlear implantee child with bilateral vestibulocochlear malformation, due to f...
متن کاملInvestigating the Effect of Cochlear Size in Insertion of Electrode Depth in Patients with Cochlear Implantation evaluated by CT-Scan
Background Cochlear implant surgery is an invasive procedure for patients with bilateral Sensorineural hearing loss and may cause some risks such as cochlear damage. We aimed to evaluate the effectiveness of cochlear measures obtained by computed tomography (CT) scan in predicting depth of cochlear implant insertion. Materials and Methods This study was conducted in a retrospective and cross-se...
متن کاملDichotic speech recognition by bilateral cochlear implant users
A series of experiments was run to assess bilateral cochlear implant users’ ability to fuse information presented dichotically. Two different methods of splitting the spectral information were investigated. In the first method, the odd-index channels were presented to one ear and the evenindex channels to the other ear. In the second method, the lower frequency channels were presented to one ea...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Annals of otology, rhinology, and laryngology
دوره 115 6 شماره
صفحات -
تاریخ انتشار 2006