Evaluation of Multiband EPI Acquisitions for Resting State fMRI
نویسندگان
چکیده
Functional magnetic resonance imaging (fMRI) and particularly resting state fMRI (rs-fMRI) is widely used to investigate resting state brain networks (RSNs) on the systems level. Echo planar imaging (EPI) is the state-of-the-art imaging technique for most fMRI studies. Therefore, improvements of EPI might lead to increased sensitivity for a large amount of studies performed every day. A number of developments to shorten acquisition time have been recently proposed and the multiband technique, allowing the simultaneous acquisition of multiple slices yielding an equivalent reduction of measurement time, is the most promising among them. While the prospect to significantly reduce acquisition time by means of high multiband acceleration factors (M) appears tempting, signal quality parameters and the sensitivity to detect common RSNs with increasing M-factor have only been partially investigated up to now. In this study, we therefore acquired rs-fMRI data from 20 healthy volunteers to systematically investigate signal characteristics and sensitivity for brain network activity in datasets with increasing M-factor, M = 2 - 4. Combined with an inplane, sensitivity encoding (SENSE), acceleration factor, S = 2, we applied a maximal acceleration factor of 8 (S2×M4). Our results suggest that an M-factor of 2 (total acceleration of 4) only causes negligible SNR decrease but reveals common RSN with increased sensitivity and stability. Further M-factor increase produced random artifacts as revealed by signal quality measures that may affect interpretation of RSNs under common scanning conditions. Given appropriate hardware, a mb-EPI sequence with a total acceleration of 4 significantly reduces overall scanning time and clearly increases sensitivity to detect common RSNs. Together, our results suggest mb-EPI at moderate acceleration factors as a novel standard for fMRI that might increase our understanding of network dynamics in healthy and diseased brains.
منابع مشابه
Detecting Resting-state Neural Connectivity Using Dynamic Network Analysis on Multiband fMRI Data
This paper describes an approach of using dynamic Structural Equation Modeling (SEM) analysis to estimate the connectivity networks from resting-state fMRI data measured by a multiband EPI sequence. Two structural equation models were estimated at each voxel with respect to the sensory-motor network and defaultmode network. The resulting connectivity maps indicate that supplementary motor area ...
متن کاملThree dimensional echo-planar imaging at 7 Tesla
Functional MRI (fMRI) most commonly employs 2D echo-planar imaging (EPI). The advantages for fMRI brought about by the increasingly popular ultra-high field strengths are best exploited in high-resolution acquisitions, but here 2D EPI becomes impractical for several reasons, including the very long volume acquisitions times. In this study at 7 T, a 3D EPI sequence with full parallel and partial...
متن کاملInvestigation of True High Frequency Electrical Substrates of fMRI-Based Resting State Networks Using Parallel Independent Component Analysis of Simultaneous EEG/fMRI Data
Previous work using simultaneously acquired electroencephalography (EEG) and functional magnetic resonance imaging (fMRI) data has shown that the slow temporal dynamics of resting state brain networks (RSNs), e.g., default mode network (DMN), visual network (VN), obtained from fMRI are correlated with smoothed and down sampled versions of various EEG features such as microstates and band-limite...
متن کاملEffect of Phase-Encoding Reduction on Geometric Distortion and BOLD Signal Changes in fMRI
Introduction Echo-planar imaging (EPI) is a group of fast data acquisition methods commonly used in fMRI studies. It acquires multiple image lines in k-space after a single excitation, which leads to a very short scan time. A well-known problem with EPI is that it is more sensitive to distortions due to the used encoding scheme. Source of distortion is inhomogeneity in the static B0 field that ...
متن کاملInfluence of physiological noise on accelerated 2D and 3D resting state functional MRI data at 7 T.
PURPOSE Physiological noise often dominates the blood-oxygen level-dependent (BOLD) signal fluctuations in high-field functional MRI (fMRI) data. Therefore, to optimize fMRI protocols, it becomes crucial to investigate how physiological signal fluctuations impact various acquisition and reconstruction schemes at different acquisition speeds. In particular, further differences can arise between ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 10 شماره
صفحات -