Prosthetic heart valves: evaluation of magnetic field interactions, heating, and artifacts at 1.5 T.
نویسندگان
چکیده
The purpose of this study was to use ex vivo testing techniques to determine the magnetic resonance imaging (MRI) safety aspects for 32 different heart valve prostheses that had not been evaluated previously in association with the 1.5-T MR environment. Ex vivo testing was performed using previously described techniques for the evaluation of magnetic field interactions (deflection angle and torque), heating [gel-filled phantom and fluoroptic thermometry; 15 minutes of MRI at a specific absorption rate (SAR) of 1.1 W/kg], and artifacts (using gradient echo and T1-weighted spin-echo pulse sequences). Thirteen heart valve prostheses displayed interactions with the magnetic field. However, these magnetic field interactions were considered relatively minor. Heating was < or =0.8 degrees C for these implants. Artifacts varied from mild to severe depending on the amount and type of metal used to make the particular heart valve prosthesis. For these 32 different heart valve prostheses, the relative lack of substantial magnetic field interactions and negligible heating indicate that MR procedures may be conducted safely in individuals with these implants using MR systems with static magnetic fields of 1.5 T or less.
منابع مشابه
Prosthetic heart valves and annuloplasty rings: assessment of magnetic field interactions, heating, and artifacts at 1.5 Tesla.
The purpose of this study was to determine the magnetic resonance (MR) safety aspects and artifacts for three different heart valve prostheses and two different annuloplasty rings that have not been evaluated previously in association with the 1.5-T MR environment. Ex vivo testing was performed using previously described techniques for the evaluation of magnetic field interactions (deflection a...
متن کاملAssessment of MRI issues for the Argus II retinal prosthesis.
OBJECTIVE The objective was to evaluate magnetic resonance imaging (MRI) issues (magnetic field interactions, heating, artifacts and functional alterations) at 1.5 T and 3 T for the Argus II Retinal Prosthesis (Second Sight Medical Products, Sylmar, CA, USA). MATERIALS AND METHODS Standardized protocols were used to assess magnetic field interactions (translational attraction and torque; 3 T,...
متن کاملCardiac pacemaker: in vitro assessment at 1.5 T.
BACKGROUND In vitro testing is used to determine safe parameters before performing magnetic resonance imaging (MRI) on a patient with an implant. Therefore, the objective of this study was to evaluate a cardiac pacemaker using a 1.5-T magnetic resonance (MR) system. METHODS A modern cardiac pacemaker (INSIGNIA I PLUS, Model 1298, and FINELINE II, Model 4471, pacing leads; Guidant Corporation,...
متن کاملPregnancy Outcome with Intracytoplasmic Sperm Injection Method in a Woman with Prosthetic Heart Valves
There is an increased risk of thromboembolism, anticoagulant-related hemorrhage, fetal-wastage and congestive cardiac failure in pregnant women with mechanical heart valves. In order to have a good outcome, the care of such patients must necessarily be multidisciplinary and in a well- equipped centre with adequate support services .One such patient who had mechanical mitral and aortic valves re...
متن کاملDetachable coil for cerebral aneurysms: in vitro evaluation of magnetic field interactions, heating, and artifacts at 3T.
A detachable coil was assessed for magnetic field interactions, heating (body RF coil, SAR of 2.0-W/kg), and artifacts at 3-Tesla. The coil showed no magnetic field interactions and heating was negligible (+0.3 degrees C). Therefore, this coil is safe for a patient undergoing MR imaging at 3T or less. While artifacts may impair the ability to properly visualize anatomy in proximity to this impl...