The multilayered film analyzer: glucose in serum, plasma, cerebrospinal fluid, and urine; and urea nitrogen in serum and plasma.

نویسندگان

  • K Warren
  • N P Kubasik
  • B B Brody
  • H E Sine
  • J P D'Souza
چکیده

Some basic laboratory performance criteria were studied for the Eastman Kodak glucose and urea nitrogen analyzer. Serum, plasma, urine, and cerebrospinal fluid were tested. Precision, both "within-day" and "between-day," for both analytes was less than 2.2% (CV). Both analytes demonstrated linearity, with about 100% recovery of added substrates. Comparisons with continuous-flow procedures demonstrated good correlation. A variety of sera and plasmas can be used for glucose analysis; oxalate--fluoride-treated plasma is unacceptable for use in analysis for urea nitrogen. We saw no effects on glucose or urea nitrogen analysis from hemolysis, lipemia, icterus, some common drugs, ammonia, or abnormally high protein concentrations. Minimum amounts detectable were: glucose (serum) 104 mg/L; glucose (cerebrospinal fluid) 100 mg/L; and urea nitrogen (serum) 21 mg/L. Calibration procedures are discussed for protein-based and aqueous samples.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Applicability of the Ektachem 400 analyzer for assaying analytes in miscellaneous body fluids.

The Ektachem analyzer, a slide-film chemistry technology, demonstrated applicability to non-serum, non-urine stimulated and actual clinical specimens having a wide range of protein concentration and pH. The simulated specimens studied had pH and protein that ranged from 2 to 9 and 0 to 88 g/L, respectively. Actual patients' specimens studied, which had protein concentrations ranging from 0.1 to...

متن کامل

Evaluation of the effect of exercise on serum creatinine and urea nitrogen in hemodialysis patients

Introduction. Chronic renal failure is a progressive disease in which kidney function is impaired, so that the body can not metabolize and maintain fluid balance and electrolytes. Therefore, the present study investigated changes in serum creatinine and urea nitrogen in hemodialysis patients in response to exercise. Materials and Methods. The subjects of this study were dialysis patients who we...

متن کامل

Computation of In Vivo Antidiabetic Activity of Holarrhena Antidysenterica Seeds Extracts in Streptozotocin-Induced Diabetic Rats

Medicinal plants have curative properties due to the presence of various complex chemical compositions, which are found as secondary plant metabolites in one or more parts of the plant. The aim of this study was to compute the antidiabetic activity of Holarrhena antidysenterica   seeds extract in streptozotocin-induced diabetic rats. The experimental protocol designed as animals were divided i...

متن کامل

Rate of Glucose Utilization by Blood Cells in Serum and Plasma Specimens With or Without Using Preservative

Objective: This study aims to estimate and compare the time-course change in blood glucose levels by blood cells in serum, and plasma with or without preservatives, which may reflect the rate of glucose utilization by blood cells. Method: This laboratory-based cross-sectional study was carried out using a blood specimen of 28 participants among which 14 were diagnosed with diabetes and 14 were...

متن کامل

Piccolo® Comprehensive Metabolic Reagent Disc

chemistry analyzer, is intended to be used for the in vitro quantitative determination of alanine aminotransferase (ALT), albumin, alkaline phosphatase (ALP), aspartate aminotransferase (AST), calcium, chloride, creatinine, glucose, potassium, sodium, total bilirubin, total carbon dioxide, total protein, and blood urea nitrogen (BUN) in heparinized whole blood, heparinized plasma, or serum in a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 26 1  شماره 

صفحات  -

تاریخ انتشار 1980