Performance of ICU ventilators during noninvasive ventilation with large leaks in a total face mask: a bench study* **

نویسندگان

  • Maria Aparecida Miyuki Nakamura
  • Eduardo Leite Vieira Costa
  • Carlos Roberto Ribeiro Carvalho
  • Mauro Roberto Tucci
چکیده

OBJECTIVE Discomfort and noncompliance with noninvasive ventilation (NIV) interfaces are obstacles to NIV success. Total face masks (TFMs) are considered to be a very comfortable NIV interface. However, due to their large internal volume and consequent increased CO2 rebreathing, their orifices allow proximal leaks to enhance CO2 elimination. The ventilators used in the ICU might not adequately compensate for such leakage. In this study, we attempted to determine whether ICU ventilators in NIV mode are suitable for use with a leaky TFM. METHODS This was a bench study carried out in a university research laboratory. Eight ICU ventilators equipped with NIV mode and one NIV ventilator were connected to a TFM with major leaks. All were tested at two positive end-expiratory pressure (PEEP) levels and three pressure support levels. The variables analyzed were ventilation trigger, cycling off, total leak, and pressurization. RESULTS Of the eight ICU ventilators tested, four did not work (autotriggering or inappropriate turning off due to misdetection of disconnection); three worked with some problems (low PEEP or high cycling delay); and one worked properly. CONCLUSIONS The majority of the ICU ventilators tested were not suitable for NIV with a leaky TFM.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Neonatal and adult ICU ventilators to provide ventilation in neonates, infants, and children: a bench model study.

BACKGROUND Using a bench test model, we investigated the hypothesis that neonatal and/or adult ventilators equipped with neonatal/pediatric modes currently do not reliably administer pressure support (PS) in neonatal or pediatric patient groups in either the absence or presence of air leaks. METHODS PS was evaluated in 4 neonatal and 6 adult ventilators using a bench model to evaluate trigger...

متن کامل

A bench study of the effects of leak on ventilator performance during noninvasive ventilation.

BACKGROUND During noninvasive ventilation (NIV), leak around the mask may cause inadequate ventilatory support or patient-ventilator asynchrony such as auto-triggering. Some NIV ventilators may be better than others at compensating for leak. METHODS We bench studied 3 NIV ventilators (Vision, Carina, and Trilogy100) and 2 ICU ventilators (Puritan Bennett 840 and Evita XL) to assess how they ...

متن کامل

Varying leaks: a challenge for modern ventilators?

Using patient-triggered ventilation, during which the patient is given the opportunity to trigger and cycle the ventilator, allows reducing the use of sedation,1 reduces ventilator-induced diaphragmatic dysfunction,2 facilitates weaning from mechanical ventilation,3 and improves patient comfort.4 To be efficient and comfortable, patient-triggered modes have to deliver pressurization in synchron...

متن کامل

Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation.

OBJECTIVES During noninvasive positive pressure ventilation (NPPV), the gas leak that commonly occurs around the mask can render NPPV ineffective. We evaluated the effects of gas leak on inspiratory trigger function during NPPV with bilevel pressure and ICU ventilators. In addition, we evaluated the effects of gas leak on fraction of inspired oxygen (Fio(2)) and humidification. METHODS Air le...

متن کامل

Influence of dynamic leaks in volume-targeted pressure support noninvasive ventilation: a bench study.

INTRODUCTION The effect of leaks on volume-targeted pressure support noninvasive ventilation mode has only been tested with continuous simulated leaks. The objective of the study was to assess the influence of random leaks occurring either during inspiration or expiration. METHODS Analysis of the volume-targeted pressure support mode in 6 commercial ventilators with single-limb circuits and i...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 40  شماره 

صفحات  -

تاریخ انتشار 2014