Study of parameters for evaluating flow reduction with stents in a sidewall aneurysm phantom model.

نویسندگان

  • Chang Ho Yu
  • Tae Kyu Kwon
چکیده

The effect of stent design parameters such as porosity, pore density, number of strands, and strut angle to the artery were studied in vitro using particle image velocimetry (PIV). Five mesh stents were implanted into a sidewall aneurysm model. The flow features in a sidewall aneurysm silicone phantom model were investigated at a Reynolds number of 300. It was found that the lowest porosity stent had the best value for velocity and vorticity reduction in an aneurysm pocket. The stent with higher pore density had a tendency to decrease the mean and maximum velocities, but it was not superior to the effects of porosity. In addition, investigation of the evaluation system confirmed that higher stent strut angles to the parent artery had a tendency to lower mean velocity, as shown by PIV and CFD. However, this effect was relatively smaller compared to porosity and pore density. Our evaluation system suggested the best combinations of parameters for the development of an ideal stent would be lower porosity, higher pore density, and higher strut angle. The results obtained in this study indicated that our evaluation system could be useful for various purposes related to evaluation of endovascular interventional devices.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Flow changes caused by the sequential placement of stents across the neck of sidewall cerebral aneurysms.

OBJECT The goal of this study was to quantify the reduction in velocity, vorticity, and shear stresses resulting from the sequential placement of stents across the neck of sidewall cerebral aneurysms. METHODS A digital particle image velocimetry (DPIV) system was used to measure the pulsatile velocity field within a flexible silicone sidewall intracranial aneurysm model and at the aneurysm ne...

متن کامل

Particle Imaging Velocimetry Evaluation of Intracranial Stents in Sidewall Aneurysm: Hemodynamic Transition Related to the Stent Design

We investigated the flow modifications induced by a large panel of commercial-off-the-shelf (COTS) intracranial stents in an idealized sidewall intracranial aneurysm (IA). Flow velocities in IA silicone model were assessed with and without stent implantation using particle imaging velocimetry (PIV). The use of the recently developed multi-time-lag method has allowed for uniform and precise meas...

متن کامل

Blood flow in intracranial aneurysms treated with Pipeline embolization devices: computational simulation and verification with Doppler ultrasonography on phantom models

PURPOSE The aim of this study was to validate a computational fluid dynamics (CFD) simulation of flow-diverter treatment through Doppler ultrasonography measurements in patient-specific models of intracranial bifurcation and side-wall aneurysms. METHODS Computational and physical models of patient-specific bifurcation and sidewall aneurysms were constructed from computed tomography angiograph...

متن کامل

Figure 1: Schematic of the Idealized Sidewall Aneurysm Model (left) and Camera and Fluoroscopic Shots of the Five Devices Implanted in Silicone Models

Brain aneurysms are abnormal ballooning of intracranial arteries which when left untreated may result in fatal outcome for the patients. Neurovascular stents have been used extensively in stent-assisted coiling treatment of aneurysms. Treatment with flow diverters (low porosity, fine-mesh stents) is used for complicated aneurysm geometries not amenable to coiling. Angiographic analysis with mat...

متن کامل

In vitro and in silico study of intracranial stent treatments for cerebral aneurysms: effects on perforating vessel flows.

BACKGROUND Many cerebral aneurysms can be treated effectively with intracranial stents. Unfortunately, stents can occlude perforating vessels near the treatment site which can decrease cerebral perfusion and increase the risk of stroke. METHODS Particle image velocimetry was used to investigate the effects of intracranial stents on flows in perforators near a treated aneurysm. In Phase 1 of t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Bio-medical materials and engineering

دوره 24 6  شماره 

صفحات  -

تاریخ انتشار 2014